Examination of International Normalised Ratio (INR) imprecision by comparison of exact and approximate formulas. Intermountain Laboratory Data Project.

S T Bennett1, G C Critchfield.   

Abstract

AIM: To evaluate the accuracy of the approximate linear formula of International Normalised Ratio (INR) imprecision by formal mathematical analysis.
METHODS: Using probability theory, an exact formula for the coefficient of variation (CV) of the INR was derived. The CV from the approximate formula was compared with the CV from the exact formula for INR determinations between 1.0 and 10.0 with International Sensitivity Indices (ISIs) between 1.0 and 3.0 and prothrombin time ratio CVs between 1.0 and 10.0%.
RESULTS: When the ISI equals 1.0, the approximate formula and the exact formula are equal. When the ISI is more than 1.0, the approximate formula overestimates the exact CV, but by less than one hundredth of the exact CV in the parameter ranges studied. The approximate formula is most accurate when laboratories achieve excellent prothrombin time measurement precision and use sensitive thromboplastins.
CONCLUSIONS: The approximate formula provides a simple means for estimating the imprecision of the INR and is sufficiently accurate to warrant its use in clinical laboratories.

Mesh:

Year:  1994        PMID: 8089220      PMCID: PMC502094          DOI: 10.1136/jcp.47.7.635

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  5 in total

Review 1.  Oral anticoagulants. Mechanism of action, clinical effectiveness, and optimal therapeutic range.

Authors:  J Hirsh; J E Dalen; D Deykin; L Poller
Journal:  Chest       Date:  1992-10       Impact factor: 9.410

2.  Effect of international sensitivity index (ISI) of thromboplastins on precision of international normalised ratios (INR)

Authors:  D A Taberner; L Poller; J M Thomson; K V Darby
Journal:  J Clin Pathol       Date:  1989-01       Impact factor: 3.411

3.  Progress in laboratory control of oral anticoagulants.

Authors:  E A Loeliger; S M Lewis
Journal:  Lancet       Date:  1982-08-07       Impact factor: 79.321

4.  The international normalized ratio and uncertainty. Validation of a probabilistic model.

Authors:  G C Critchfield; S T Bennett
Journal:  Am J Clin Pathol       Date:  1994-07       Impact factor: 2.493

5.  Effect of laboratory variation in the prothrombin-time ratio on the results of oral anticoagulant therapy.

Authors:  M H Eckman; H J Levine; S G Pauker
Journal:  N Engl J Med       Date:  1993-09-02       Impact factor: 91.245

  5 in total
  1 in total

Review 1.  Uncertainty in measurement: A review of the procedures for determining uncertainty in measurement and its use in deriving the biological variation of the estimated glomerular filtration rate.

Authors:  Ian Farrance; Tony Badrick; Robert Frenkel
Journal:  Pract Lab Med       Date:  2018-04-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.