Use of polyacrylamide gels in a saline-filled tank to determine the linearity of the Sheffield Mark 1 electrical impedance tomography (EIT) system in measuring impedance disturbances.

D S Holder1, A Khan.   

Abstract

Polyacrylamide gels have been assessed as a means of producing reliable small impedance changes for EIT studies in saline-filled tanks. Test rods could be made in about 10 min from a stock solution of acrylamide monomer and bis-acrylamide, ammonium persulphate and tetramethylethylaminediamine, completed with saline. Impedance increases of about 10%-200% were produced by varying the gel concentration. The method appears to be easy to use, inexpensive and is stable over time. Its use was illustrated by acquiring EIT images with a Sheffield Mark 1 system and a tank filled with 0.2% saline. Polythene or polyacrylamide rods of varying resistivities, 10% of the tank diameter wide, were placed in one of three positions. The magnitude of the impedance increase in EIT images varied linearly with the actual impedance change in the rods for increases up to 20%, but the blurring in images was similar when polythene rods were used. This suggests that the use of highly resistive test objects gives a reasonable indication of the blurring in EIT images acquired with the system used, but measurement of the magnitude of the impedance change is more likely to be accurate for changes of a few tens of per cent.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8087049     DOI: 10.1088/0967-3334/15/2a/006

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  3 in total

Review 1.  [Electrical impedance tomography: ready for routine clinical use for mechanically ventilated patients?].

Authors:  J Hinz; G Hahn; M Quintel
Journal:  Anaesthesist       Date:  2008-01       Impact factor: 1.041

2.  A versatile high-permittivity phantom for EIT.

Authors:  Tzu-Jen Kao; Gary J Saulnier; David Isaacson; Tomas L Szabo; Jonathan C Newell
Journal:  IEEE Trans Biomed Eng       Date:  2008-11       Impact factor: 4.538

3.  Fast detection and data compensation for electrodes disconnection in long-term monitoring of dynamic brain electrical impedance tomography.

Authors:  Ge Zhang; Meng Dai; Lin Yang; Weichen Li; Haoting Li; Canhua Xu; Xuetao Shi; Xiuzhen Dong; Feng Fu
Journal:  Biomed Eng Online       Date:  2017-01-07       Impact factor: 2.819

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.