Outbreak of pyrogenic reactions and gram-negative bacteremia in a hemodialysis center.

B M Jackson1, C M Beck-Sague, L A Bland, M J Arduino, L Meyer, W R Jarvis.   

Abstract

Six episodes of gram-negative bacteremia and seven pyrogenic reactions occurred in 11 patients in one hemodialysis center. Gram-negative bacteremias and/or pyrogenic reactions were not related to reuse and were more likely to occur if dialysis was performed in one unit of the center (8/13 unit 5 vs. 221/1,151 in other units, p < 0.001) and with one type of dialysis machine (10/13 vs. 581/1,151 with other machines, p = 0.05), which was preferentially used in unit 5 (p < 0.01). Bacterial and endotoxin concentrations of water used to prepare dialysate and reprocess hemodialyzers, and of dialysate, exceeded allowable concentrations recommended by the Association for the Advancement of Medical Instrumentation (AAMI). The implicated dialysis machines were disinfected with chemicals daily, but not heat-disinfected daily as suggested by the manufacturer. Results suggest that the outbreak was caused by the use of water that did not meet AAMI standards and inadequate disinfection of one type of dialysis machine.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8080011     DOI: 10.1159/000168694

Source DB:  PubMed          Journal:  Am J Nephrol        ISSN: 0250-8095            Impact factor:   3.754


  4 in total

1.  Detection of endotoxin using an evanescent wave fiber-optic biosensor.

Authors:  E A James; K Schmeltzer; F S Ligler
Journal:  Appl Biochem Biotechnol       Date:  1996-09       Impact factor: 2.926

Review 2.  Hemodialysis and water quality.

Authors:  Angela D Coulliette; Matthew J Arduino
Journal:  Semin Dial       Date:  2013 Jul-Aug       Impact factor: 3.455

3.  Microbiological qualification of air, water and dialysate in a haemodialysis centre: a new focus on Legionella spp.

Authors:  Saeid Nazemi; Mehdi Mirzaii; Somayeh Yaslianifard; Davood Darban-Sarokhalil; Seyyed Sajjad Khoramrooz; Pirasteh Norozi; Fatemeh Davardoost
Journal:  Iran J Microbiol       Date:  2016-08

4.  Prevent infection linked to the dialysis water in a hemodialysis center in Fez city (Morocco).

Authors:  Bouchra Oumokhtar; Abdelhakim El Ouali Lalami; Mustapha Mahmoud; Sanae Berrada; Mohammed Arrayhani; Tarik Squalli Houssaini
Journal:  Pan Afr Med J       Date:  2013-11-28
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.