Literature DB >> 807229

A critical look at chemical disinfection of anaesthetic apparatus.

R H George.   

Abstract

Studies were carried out to test the efficacy of disinfection of a previously contaminated corrugated anaesthetic tube and reservoir bag. Bacteriological swabs and sampling with broth were evaluated as methods of detecting contamination. Gross contamination was detected by broth sampling on occasions when swabbing failed to reveal contamination. The presence of up to 18% by volume of residual air within the apparatus was detected. A method of ensuring elimination of all air is described and studies showed that disinfection was then complete.

Mesh:

Substances:

Year:  1975        PMID: 807229     DOI: 10.1093/bja/47.6.719

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166