Literature DB >> 8065694

Suction attachment for stereotactic radiosurgery of intraocular malignancies.

M Zehetmayer1, R Menapace, K Kitz, A Ertl.   

Abstract

We designed a suction attachment for the radiosurgical treatment of intraocular malignancies with the Leksell gamma unit (Gamma Knife). Our device consists of a circular suction chamber and an adjustable unit to be fixed to the Leksell stereotactic head frame. All components are made of plastic materials in order to avoid artifacts in CT or MRT imaging. A permanent suction of 600-800 mbar is provided by a standard vacuum pump, powered by a portable battery. Suction times up to 40 min were well tolerated in all cases. We successfully used this device, performing 17 radiosurgical treatments in 8 patients with large or extra-large uveal melanomas and one patient suffering from a choroidal metastasis.

Entities:  

Mesh:

Year:  1994        PMID: 8065694     DOI: 10.1159/000310467

Source DB:  PubMed          Journal:  Ophthalmologica        ISSN: 0030-3755            Impact factor:   3.250


  3 in total

1.  Radiotherapy in ophthalmology : 2004 Jules Gonin lecture of the Retina Research Foundation.

Authors:  Leonidas Zografos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-07       Impact factor: 3.117

2.  Ocular arterial blood flow of choroidal melanoma eyes before and after stereotactic radiotherapy using Leksell gamma knife: 2 year follow up.

Authors:  P V Vécsei; K Kircher; G Nagel; S Toma-Bstaendig; I Ruhswurm; M Georgopoulos; M Zehetmayer
Journal:  Br J Ophthalmol       Date:  1999-12       Impact factor: 4.638

3.  Gamma Knife Perfexion® radiosurgery and endo diode laser thermotherapy for choroidal melanoma with technical analysis: A case report.

Authors:  Yi-Chieh Tsai; Chun-Yuan Kuo; Jia-Wei Lin; Shun-Tai Yang; Shih-Chung Lai; Jo-Ting Tsai
Journal:  Oncol Lett       Date:  2017-10-31       Impact factor: 2.967

  3 in total