Design and test of a hands-free tracheostoma valve to improve the rehabilitation process after laryngectomy.

G J Verkerke1, A Veenstra, H K Schutte, I F Herrmann, G Rakhorst.   

Abstract

The surgical treatment of throat cancer often requires total laryngectomy. The necessary tracheostoma attracts attention, especially during speech, when one must close it manually to force air into the esophagus. To avoid this manual control, several devices have been developed. These must be placed in or around the tracheostoma and possess a valve which closes with a large air flow value. Most devices have a major drawback; coughing calls for manual adaptation, which again attracts attention. A new hands-free tracheostoma valve has been developed. The device consists of two magnetic valves, one which closes the outflow to allow speech, and one which opens to cough. Prototypes were tested in a physical model. Air flow resistance and closing pressures of the speech valve proved to be comparable to those of existing devices. Patients were enthusiastic about the new approach, even though air leakage sometimes occurred.

Entities:  

Mesh:

Year:  1994        PMID: 8050810

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  2 in total

1.  Hands-free speech after surgical voice rehabilitation with a Provox voice prosthesis: experience with the Provox FreeHands HME tracheostoma valve system.

Authors:  K J Lorenz; K Groll; A H Ackerstaff; F J M Hilgers; H Maier
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-10-17       Impact factor: 2.503

2.  [Finger-free speech with the Provox FreeHands HME Automatic Tracheostoma Valve system. Clinical long-term experience].

Authors:  K J Lorenz; K Huverstuhl; H Maier
Journal:  HNO       Date:  2009-11       Impact factor: 1.284

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.