K Miyajima1, M Hirata, T Yoshida, H Kosaka, A Okayama.
Abstract
A method for the determination of delta-aminolevulinic acid in plasma of lead-exposed workers by high-performance liquid chromatography with fluorescence detection of a fluorescent delta-aminolevulinic acid derivative (2-methylidineamino-3,5-diacetyl-4,6-dimethylpropionic acid) was established. The detection limit of delta-aminolevulinic acid in plasma was 0.01 microgram/ml at a signal-to-noise ratio of 5:1. A linear correlation was obtained between the amounts of delta-aminolevulinic acid injected from 0.01 to 0.5 microgram/ml (r = 0.999). The recovery of 0.05 and 0.1 microgram/ml of delta-aminolevulinic acid added to plasma with various concentrations of delta-aminolevulinic acid in plasma ranged from 80.0 to 100.8%. This method, combined with the use of an automatic sampler, should facilitate the routine measurement of delta-aminolevulinic acid in plasma.Entities:
Mesh:
Substances:
Year: 1994 PMID: 8044276 DOI: 10.1016/0378-4347(94)00046-8