A comparative study of the fetal electrocardiogram recorded by the STAN an Nottingham systems.

L Skillern1, J Cockburn, M Benjamin, J M Pearce, D Sahota, N Reed, M Mohajer, D James, M Symonds.   

Abstract

OBJECTIVE: To compare the T:QRS ratio recorded by the STAN and Nottingham fetal electrocardiogram (FECG) monitors.
DESIGN: Prospective observational study.
SETTING: London teaching hospital delivery suite and research unit. METHODS AND
SUBJECTS: The T:QRS ratios generated by the STAN and Nottingham FECG monitors were simultaneously recorded and compared using signals generated from a computer-produced ECG signal and signals from 11 term fetuses recorded during labour.
RESULTS: There was an acceptable level of agreement between the two systems with the computer-generated signals, but it was not clinically acceptable with the signals from the fetuses recorded during labour. Disagreements in the T:QRS values were probably due to differences in the reference points for the measurement of the S-T segment and T-wave height.
CONCLUSION: The different points of reference for measurement of S-T segment and T-wave height can explain poor agreement between the two methods of FECG waveform analysis. The suggested adopted points of reference are those corresponding to adult electrocardiographic methodology.

Mesh:

Year:  1994        PMID: 8043535     DOI: 10.1111/j.1471-0528.1994.tb13647.x

Source DB:  PubMed          Journal:  Br J Obstet Gynaecol        ISSN: 0306-5456