Three-dimensional echocardiography: in vitro and in vivo validation of left ventricular mass and comparison with conventional echocardiographic methods.

A S Gopal1, A M Keller, Z Shen, P M Sapin, K M Schroeder, D L King, D L King.   

Abstract

OBJECTIVES: This study aimed to validate a method for mass computation in vitro and in vivo and to compare it with conventional methods.
BACKGROUND: Conventional echocardiographic methods of determining left ventricular mass are limited by assumptions of ventricular geometry and image plane positioning. To improve accuracy, we developed a three-dimensional echocardiographic method that uses nonparallel, nonintersecting short-axis planes and a polyhedral surface reconstruction algorithm for mass computation.
METHODS: Eleven fixed hearts were imaged by three-dimensional echocardiography, and mass was determined in vitro by multiplying the myocardial volume by the density of each heart and comparing it with the true mass. Mass at diastole and systole by three-dimensional echocardiography and magnetic resonance imaging (MRI) was compared in vivo in 15 normal subjects. Ten subjects also underwent imaging by one- and two-dimensional echocardiography, and mass was determined by Penn convention, area-length and truncated ellipsoid algorithms.
RESULTS: In vitro results were r = 0.995, SEE 2.91 g, accuracy 3.47%. In vivo interobserver variability for systole and diastole was 16.7% to 27%, 14% to 18.1% and 6.3% to 12.8%, respectively, for one-, two- and three-dimensional echocardiography and was 7.5% for MRI at end-diastole. The latter two agreed closely with regard to diastolic mass (r = 0.895, SEE 11.1 g) and systolic mass (r = 0.926, SEE 9.2 g). These results were significantly better than correlations between MRI and the Penn convention (r = 0.725, SEE 25.6 g for diastole; r = 0.788, SEE 28.7 g for systole), area-length (r = 0.694, SEE 24.2 g for diastole; r = 0.717, SEE 28.2 g for systole) and truncated ellipsoid algorithms (r = 0.687, SEE 21.8 g for diastole; r = 0.710, SEE 24.5 g for systole).
CONCLUSIONS: Image plane positioning guidance and elimination of geometric assumptions by three-dimensional echocardiography achieve high accuracy for left ventricular mass determination in vitro. It is associated with higher correlations and lower standard errors than conventional methods in vivo.

Mesh:

Year:  1994        PMID: 8034889     DOI: 10.1016/0735-1097(94)90310-7

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  17 in total

Review 1.  Developments in cardiovascular ultrasound. Part 3: Cardiac applications.

Authors:  C M Moran; W N McDicken; P R Hoskins; P J Fish
Journal:  Med Biol Eng Comput       Date:  1998-09       Impact factor: 2.602

2.  Elongation as a new shape index for the left ventricle.

Authors:  B Knap; G Juznic; A F Bren; G Drzewiecki; A Noordergraaf
Journal:  Int J Cardiovasc Imaging       Date:  2002-12       Impact factor: 2.357

3.  Determination of left ventricular mass by three-dimensional echocardiography: in vitro validation of a novel quantification method using multiple equi-angular rotational planes for rapid measurements.

Authors:  Claudius Teupe; Masaaki Takeuchi; Jiefen Yao; Natesa Pandian
Journal:  Int J Cardiovasc Imaging       Date:  2002-06       Impact factor: 2.357

4.  Three-dimensional echocardiographic measurement of left and right ventricular mass and volume: in vitro validation.

Authors:  Michal Hubka; Edward L Bolson; John A McDonald; Roy W Martin; Brad Munt; Florence H Sheehan
Journal:  Int J Cardiovasc Imaging       Date:  2002-04       Impact factor: 2.357

Review 5.  Left ventricular assessment using real time three dimensional echocardiography.

Authors:  L Sugeng; L Weinert; R M Lang
Journal:  Heart       Date:  2003-11       Impact factor: 5.994

6.  Improved quantification of left ventricular mass based on endocardial and epicardial surface detection with real time three dimensional echocardiography.

Authors:  E G Caiani; C Corsi; L Sugeng; P MacEneaney; L Weinert; V Mor-Avi; R M Lang
Journal:  Heart       Date:  2005-05-12       Impact factor: 5.994

7.  Clinical application of three-dimensional echocardiography: past, present and future.

Authors:  S A Kleijn; O Kamp
Journal:  Neth Heart J       Date:  2009-01       Impact factor: 2.380

Review 8.  Three-dimensional adult echocardiography: where the hidden dimension helps.

Authors:  Victor Mor-Avi; Lissa Sugeng; Roberto M Lang
Journal:  Curr Cardiol Rep       Date:  2008-05       Impact factor: 2.931

9.  Treating anemia in older adults with heart failure with a preserved ejection fraction with epoetin alfa: single-blind randomized clinical trial of safety and efficacy.

Authors:  Mathew S Maurer; Sergio Teruya; Bibhas Chakraborty; Stephen Helmke; Donna Mancini
Journal:  Circ Heart Fail       Date:  2012-12-20       Impact factor: 8.790

10.  Left ventricular mass in 169 healthy children and young adults assessed by three-dimensional echocardiography.

Authors:  T Poutanen; E Jokinen
Journal:  Pediatr Cardiol       Date:  2007-05-05       Impact factor: 1.655

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.