Motion and laxity of the capitellocondylar total elbow prosthesis.

G J King1, E Itoi, G L Niebur, B F Morrey, K N An.   

Abstract

The motion and laxity of the capitellocondylar unconstrained total elbow prosthesis were assessed, with use of an electromagnetic tracking device and stimulated muscle-loading, after implantation in seventeen cadaveric elbows. The axis of motion of the elbows with the capitellocondylar implants averaged 2.1 +/- 2.3 degrees more varus angulation than that of the intact elbows. This difference may be attributed to the design of the implant, as the 5-degree-valgus humeral component used in this study has a smaller valgus inclination than the articular surface of the distal aspect of the humerus. Although the maximum valgus-varus laxity of the capitellocondylar elbow prostheses was, on the average, 4.3 +/- 2.4 degrees greater than normal (with simulated muscle-loading), the data must be interpreted in light of the fact that this in vitro study did not allow for soft-tissue healing. The prosthetic components tracked well, and there were no dislocations or malarticulations provided that appropriate soft-tissue tensioning and positioning of the components had been achieved at the time of implantation. Sectioning of either the medial or the lateral collateral ligament resulted in gross instability of the joint after capitellocondylar arthroplasty. The ulnar attachment of the medial collateral ligament was found to be vulnerable to injury during the positioning of the ulnar component of this implant.

Entities:  

Mesh:

Year:  1994        PMID: 8027106     DOI: 10.2106/00004623-199407000-00007

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  4 in total

Review 1.  [Revision arthroplasty of the elbow].

Authors:  K J Burkhart; G Stein; E Skouras; L P Müller
Journal:  Unfallchirurg       Date:  2010-12       Impact factor: 1.000

2.  The effect of implant linking and ligament integrity on humeral loading of a convertible total elbow arthroplasty.

Authors:  James R Brownhill; J Whitcomb Pollock; Louis M Ferreira; James A Johnson; Graham Jw King
Journal:  Shoulder Elbow       Date:  2017-08-31

Review 3.  Current concepts of elbow-joint disorders and their treatment.

Authors:  Katsunori Inagaki
Journal:  J Orthop Sci       Date:  2013-01-11       Impact factor: 1.601

4.  A physiological dynamic testing machine for the elbow joint.

Authors:  Johannes Kiene; Robert Wendlandt; Marcus Heinritz; Angelika Schall; Arndt-Peter Schulz
Journal:  Open Orthop J       Date:  2013-04-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.