Accuracy of a new transmittance-reflectance pulse oximetry sensor in critically ill neonates.

S Fanconi1, A Tschupp.   

Abstract

OBJECTIVE: To test the accuracy of a new pulse oximeter sensor based on transmittance and reflectance. This sensor makes transillumination of tissue unnecessary and allows measurements on the hand, forearm, foot, and lower limb.
DESIGN: Prospective, open, nonrandomized criterion standard study.
SETTING: Neonatal intensive care unit, tertiary care center. PATIENTS: Sequential sample of 54 critically ill neonates (gestational age 27 to 42 wks; postnatal age 1 to 28 days) with arterial catheters in place.
MEASUREMENTS AND MAIN RESULTS: A total of 99 comparisons between pulse oximetry and arterial saturation were obtained. Comparison of femoral or umbilical arterial blood with transcutaneous measurements on the lower limb (n = 66) demonstrated an excellent correlation (r2 = .96). The mean difference was +1.44% +/- 3.51 (SD) % (range -11% to +8%). Comparison of the transcutaneous values with the radial artery saturation from the corresponding upper limb (n = 33) revealed a correlation coefficient of 0.94 with a mean error of +0.66% +/- 3.34% (range -6% to +7%). The mean difference between noninvasive and invasive measurements was least with the test sensor on the hand, intermediate on the calf and arm, and greatest on the foot. The mean error and its standard deviation were slightly larger for arterial saturation values < 90% than for values > or = 90%.
CONCLUSION: Accurate pulse oximetry saturation can be acquired from the hand, forearm, foot, and calf of critically ill newborns using this new sensor.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8026204     DOI: 10.1097/00003246-199407000-00016

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  1 in total

1.  Individualised pulse oximetry limits in neonatal intensive care.

Authors:  R Gupta; C W Yoxall; N Subhedar; N J Shaw
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-11       Impact factor: 5.747

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.