Thermodilution right ventricular ejection fraction measurements: room temperature versus cold temperature injectate.

K Safcsak1, L D Nelson.   

Abstract

OBJECTIVE: To compare thermodilution right ventricular ejection fraction measurements using 10 mL room temperature injectate vs. 10 mL cold temperature injectate.
DESIGN: Prospective, clinical study.
SETTING: Adult surgical intensive care unit (ICU) in a university hospital. PATIENTS: Sixty adult surgical ICU patients requiring hemodynamic monitoring by a pulmonary artery catheter.
INTERVENTIONS: Patients were in a supine position with the bed flat during thermodilution measurements. Four 10 mL room temperature injections were alternated with four 10 mL cold temperature injections.
MEASUREMENTS AND MAIN RESULTS: One hundred eleven paired thermodilution right ventricular ejection fraction measurements were made in patients during a "steady state." There were no restrictions regarding body temperature, cardiac index, heart rate or rhythm. Injectate temperature was measured by an in-line temperature probe. Injections were synchronized with end-expiration of mechanical ventilator breaths. The first injection was deleted from each temperature group. Reproducibility of individual right ventricular ejection fraction measurements was assessed by calculating the mean variation of triplicate measurements in each temperature group. Mean values of room temperature measurements were compared with cold temperature measurements by Student's t-test. Linear regression analysis, bias, and precision were also calculated. There was no significant difference (p = .752) between mean right ventricular ejection fraction measurements determined with room temperature (23.9 +/- 1 degrees C) vs. cold temperature (8.0 +/- 1.1 degrees C) injectate. There was a high degree of correlation between measurements (r2 = .876, p < .001). The bias of room temperature measurements compared with cold temperature was -0.39% and the precision was +/- 3.3%. The mean variation between individual measurements in all room temperature and cold temperature right ventricular ejection fraction measurements was 9.7% and 8.0%, respectively. There was no significant difference and there was a high degree of correlation in mean right ventricular ejection fraction measurements when data were grouped according to body temperature, heart rate, cardiac index, right ventricular ejection fraction, central venous pressure, pulmonary vascular resistance index, right ventricular end-diastolic volume index, or right ventricular stroke work index.
CONCLUSIONS: The results suggest that room temperature injectate may be used for right ventricular ejection fraction measurements in critically ill adult surgical patients. Utilizing room temperature injectate for right ventricular ejection fraction measurements may save time and costs in the critical care unit.

Entities:  

Mesh:

Year:  1994        PMID: 8026203     DOI: 10.1097/00003246-199407000-00015

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  2 in total

1.  Effect of Guglielmi detachable coils on intraaneurysmal flow: experimental study in canines.

Authors:  Angelika Sorteberg; Wilhelm Sorteberg; Alan Rappe; Charles M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2002-02       Impact factor: 3.825

2.  Effects of extreme lateral posture on hemodynamics and plasma atrial natriuretic peptide levels in critically ill patients.

Authors:  T Bein; C Metz; C Keyl; M Pfeifer; K Taeger
Journal:  Intensive Care Med       Date:  1996-07       Impact factor: 17.440

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.