| Literature DB >> 8006797 |
A V Spigelman, W C Albert, C H Cozean, D G Johnson, P J McDonnell, P M Pender, J Shimmick.
Abstract
In this paper, we examine the 193 nm excimer laser's efficacy and safety in treating myopic astigmatism. The VISX Twenty/Twenty excimer laser uses aperture elements to effect astigmatic photorefractive keratectomy. In 70 patients with six months follow-up, the average postoperative sphere was -0.14 and the average postoperative cylinder was -0.54. Seventy-one percent of these patients had uncorrected visual acuity of 20/40 or better at six months. In 12 patients with one year follow-up, the average postoperative sphere was -0.05 and the average postoperative cylinder was -0.59. Eighty-three percent of these patients had uncorrected acuities of 20/40 or better; none had a clinically significant loss of best corrected acuity. This investigation demonstrates that the excimer laser can be used to treat myopic astigmatism successfully.Entities:
Mesh:
Year: 1994 PMID: 8006797 DOI: 10.1016/s0886-3350(13)80763-8
Source DB: PubMed Journal: J Cataract Refract Surg ISSN: 0886-3350 Impact factor: 3.351