| Literature DB >> 8005888 |
D P Chen1, E Jimenez, K Ataka, S Levitsky, H Feinberg.
Abstract
We describe a novel and relatively inexpensive spectrofluorescence system that supplies rapidly alternating wavelengths to either a standard cuvette or an isolated perfused heart. Its use is illustrated by determining changes in cytosolic intracellular Ca2+ concentration ([Ca2+]i) by using the Ca(2+)-sensitive fluorescent dye fura 2 in a rabbit heart preparation. The system uses two precision electromechanical shutters (capable of gating with respect to the electrocardiographic R wave for signal averaging) allowing alternate fura 2 excitation wavelengths (340 and 380 nm) without moving optical components and uses a fiber optic for conducting excitation and collecting epifluorescence. Sample recordings tracing the [Ca2+]i transient in an entire cardiac cycle and in capturing specific isolated regions (diastole and systole) of the cycle are presented. Limitations of this low-cost but easily implemented system are discussed.Entities:
Mesh:
Substances:
Year: 1994 PMID: 8005888 DOI: 10.1152/jappl.1994.76.3.1394
Source DB: PubMed Journal: J Appl Physiol (1985) ISSN: 0161-7567