Fura 2 determination of [Ca2+]i in isolated perfused heart using R wave-gated electromechanical shutters.

D P Chen1, E Jimenez, K Ataka, S Levitsky, H Feinberg.   

Abstract

We describe a novel and relatively inexpensive spectrofluorescence system that supplies rapidly alternating wavelengths to either a standard cuvette or an isolated perfused heart. Its use is illustrated by determining changes in cytosolic intracellular Ca2+ concentration ([Ca2+]i) by using the Ca(2+)-sensitive fluorescent dye fura 2 in a rabbit heart preparation. The system uses two precision electromechanical shutters (capable of gating with respect to the electrocardiographic R wave for signal averaging) allowing alternate fura 2 excitation wavelengths (340 and 380 nm) without moving optical components and uses a fiber optic for conducting excitation and collecting epifluorescence. Sample recordings tracing the [Ca2+]i transient in an entire cardiac cycle and in capturing specific isolated regions (diastole and systole) of the cycle are presented. Limitations of this low-cost but easily implemented system are discussed.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8005888     DOI: 10.1152/jappl.1994.76.3.1394

Source DB:  PubMed          Journal:  J Appl Physiol (1985)        ISSN: 0161-7567


  1 in total

1.  Cardiac electrophysiological imaging systems scalable for high-throughput drug testing.

Authors:  Peter Lee; Ken Wang; Christopher E Woods; Ping Yan; Peter Kohl; Paul Ewart; Leslie M Loew; Derek A Terrar; Christian Bollensdorff
Journal:  Pflugers Arch       Date:  2012-09-29       Impact factor: 3.657