Literature DB >> 7995745

The effectiveness of immobilization during prostate irradiation.

G C Bentel1, L B Marks, G W Sherouse, D P Spencer, M S Anscher.   

Abstract

PURPOSE: To evaluate the effect of a hemibody foam cradle on the reproducibility of patient setup during external beam radiation treatment of prostate cancer. METHODS AND MATERIALS: Between January 1992 and April 1993, 74 patients received external beam radiation treatment to the prostate +/- nodes, generally with a four-field box technique. Forty-four of the 74 patients had a custom-made hemibody foam cast used in an attempt to improve setup accuracy. A review of the routine weekly port films was performed following the completion of therapy to determine the reproducibility of patient setup in all 74 patients. The physician's request of an isocenter shift was used as an indicator of reproducibility. Neither the treating technologists nor the physicians knew at the time the films were taken that the port films would be reviewed for setup reproducibility at a later date. The results were compared between the patients treated with (44) and without (30) an immobilization device.
RESULTS: In the 44 immobilized patients, 213 routine checks of the isocenter were performed during the 7-week course of radiation therapy. In 17.4% of these instances (37 out of 213), an isocenter shift was requested. This rate is compared to 23.1% (30 out of 130) in the 30 patients who did not have the immobilization device (p < 0.2). There was a statistically significant reduction in isocenter shifts requested in the anterior to posterior direction in the patients who were immobilized, 5.1% (9 out of 175) vs. 12.6% (13 out of 103) (p < 0.05, two tailed chi-square test). There was no significant improvement in the reproducibility of isocenter placement in the cephalad to caudal or right to left directions.
CONCLUSIONS: This custom-made hemibody foam cradle appears to improve the reproducibility of patient setup during the 7-week course of fractionated external beam irradiation for patients with adenocarcinoma of the prostate. This type of immobilization device is now routinely used in our clinic and is recommended for all patients receiving pelvic radiotherapy. These devices are likely to be particularly useful when contemplating dose escalation to minimize the volume of bladder and rectum included in the treatment fields.

Entities:  

Mesh:

Year:  1995        PMID: 7995745     DOI: 10.1016/0360-3016(94)00351-K

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  8 in total

1.  Six-Dimensional Correction of Intra-Fractional Prostate Motion with CyberKnife Stereotactic Body Radiation Therapy.

Authors:  Siyuan Lei; Nathaniel Piel; Eric K Oermann; Viola Chen; Andrew W Ju; Kedar N Dahal; Heather N Hanscom; Joy S Kim; Xia Yu; Guowei Zhang; Brian T Collins; Reena Jha; Anatoly Dritschilo; Simeng Suy; Sean P Collins
Journal:  Front Oncol       Date:  2011-12-08       Impact factor: 6.244

2.  Development of patient support devices for execution of clinical radiotherapy for cancer patients: A preliminary report.

Authors:  N K Babu; Bakshish Singh; S Namrata; B K Mohanti; R Ravichandran; K E Ghamrawy
Journal:  J Med Phys       Date:  2006-10

3.  [The conformal radiotherapy of localized prostatic carcinoma: acute tolerance and early efficacy].

Authors:  D Zierhut; M Flentje; G Sroka-Perez; V Rudat; R Engenhart-Cabillic; M Wannenmacher
Journal:  Strahlenther Onkol       Date:  1997-02       Impact factor: 4.033

4.  A comparison of two systems of patient immobilization for prostate radiotherapy.

Authors:  Peter White; Chui Ka Yee; Lee Chi Shan; Lee Wai Chung; Ng Ho Man; Yik Shing Cheung
Journal:  Radiat Oncol       Date:  2014-01-22       Impact factor: 3.481

Review 5.  Image-guided radiotherapy for prostate cancer.

Authors:  Audrey Dang; Patrick A Kupelian; Minsong Cao; Nzhde Agazaryan; Amar U Kishan
Journal:  Transl Androl Urol       Date:  2018-06

6.  Verification of setup errors in external beam radiation therapy using electronic portal imaging.

Authors:  K Krishna Murthy; Zakiya Al-Rahbi; S S Sivakumar; C A Davis; R Ravichandran; Kamal El Ghamrawy
Journal:  J Med Phys       Date:  2008-04

7.  Immobilization versus no immobilization for pelvic external beam radiotherapy.

Authors:  Avinash H Udayashankar; Shibina Noorjahan; Nirmala Srikantia; K Ravindra Babu; Sandeep Muzumder
Journal:  Rep Pract Oncol Radiother       Date:  2018-05-17

8.  Accuracy and reproducibility of conformal radiotherapy using data from a randomised controlled trial of conformal radiotherapy in prostate cancer (MRC RT01, ISRCTN47772397).

Authors:  S Stanley; S Griffiths; M R Sydes; A R Moore; I Syndikus; D P Dearnaley
Journal:  Clin Oncol (R Coll Radiol)       Date:  2008-06-18       Impact factor: 4.126

  8 in total