Abstract

Polyethylene wear is currently 1 of the major problems of total knee replacement. Evaluation of a large number of revised tibial components showed 3 wear patterns. Symmetrical central and symmetrical posterior wear are rare. Asymmetric wear beginning at the posteromedial part of the tibial plateau is the common pattern. The first step in the main wear process is cold flow of the polyethylene of the posteromedial part of the tibial plateau. The horizontal axis of rotation becomes fixed by this plastic deformation, and wear of the posteromedial part of the plateau then ensues. The vertical axis of rotation also becomes trapped, resulting in external rotation of the tibia on the femur leading to maltracking, exacerbating the wear and leading to an increased volume of wear debris, which eventually leads to chronic synovitis and osteolysis. Wear is more pronounced in the thinner units. Absence of a cruciate gap seems to prevent the initial cold flow and thus minimizes wear.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7994974

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  3 in total

1.  Does total knee arthroplasty change frontal plane knee biomechanics during gait?

Authors:  Karl F Orishimo; Ian J Kremenic; Ajit J Deshmukh; Stephen J Nicholas; Jose A Rodriguez
Journal:  Clin Orthop Relat Res       Date:  2011-11-29       Impact factor: 4.176

2.  Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty.

Authors:  Kazuaki Mineta; Masahiko Okada; Soshi Matsumoto; Daisuke Hamada; Tomohiro Goto; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2017-05-16

3.  Unique relationship between osteophyte and femoral-tibia component size mismatch in determining polyethylene wear in primary total knee arthroplasty: a case report.

Authors:  Manjunath Ramappa; Andrew Port
Journal:  J Med Case Rep       Date:  2009-02-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.