Cardiac pacemaker electrodes: improved methods of extraction. Work in progress.

C W Smith1, J C Messenger, S P Schaner, J W Renner.   

Abstract

PURPOSE: To develop improved methods of extracting pacemaker electrodes from the femoral vein by using standard equipment and techniques.
MATERIALS AND METHODS: Pacemaker electrodes were removed from 61 patients aged 21-91 years (mean, 68 years). In 25 patients (41%), traction was applied to locking stylets to remove electrodes at the subclavian vein entry site. In 30 patients, the electrode tip was freed with traction at the subclavian vein entry site. The electrode was then extracted through a sheath placed in the femoral vein. In six patients (10%), the electrode tip was freed via the femoral vein when traction was applied to grasping devices. The electrode was then removed through a femoral venous sheath.
RESULTS: Ninety-nine (95%) of 104 pacemaker electrodes were extracted completely. Two electrodes were trapped in fibrous tissue in the subclavian vein. One right atrial tear occurred.
CONCLUSION: Pacemaker electrodes can be removed effectively by using percutaneous techniques.

Entities:  

Mesh:

Year:  1994        PMID: 7972816     DOI: 10.1148/radiology.193.3.7972816

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Arteriovenous fistula following pacemaker lead removal: CT diagnosis.

Authors:  Preet Kang; Diego B Nunez
Journal:  Emerg Radiol       Date:  2003-03-26

2.  Extraction and replacement of permanent pacemaker leads through occluded vessels: use of extraction sheaths as conduits--balloon venoplasty as an adjunct.

Authors:  J N Pace; M Maquilan; S E Hessen; P A Khoury; A Wilson; S P Kutalek
Journal:  J Interv Card Electrophysiol       Date:  1997-12       Impact factor: 1.900

Review 3.  Endovascular extraction techniques for pacemaker and ICD lead extraction: Part 1.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-04       Impact factor: 2.380

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.