[The Heidelberg Retina Tomograph: reproducibility and measuring errors in different pupillary widths using a model eye].

P Janknecht1, J Funk.   

Abstract

BACKGROUND: We measured the reproducibility and relative error of the Heidelberg Retina Tomograph (HRT) in a model eye with various widths of the pupil.
METHODS: We put diaphragms of various diameters in front of the lens of a model eye. The volume of an artificial papilla which was inserted into the model eye was measured three times. The coefficient of variation (standard deviation/mean value) was taken as a measure of the reproducibility. The relative error was defined as the absolute value of the quotient of the difference between measured and real value.
RESULTS: The relative error for diaphragms of 1.5 to 5 mm was 7% to 18.4%, for a diaphragm of 1 mm 19.8%, and for a diaphragm of 11 mm 21.3%. The coefficient of variation ranged from 0.49% to 1.95% for diaphragms of 1.5 to 11 mm diameter, and was 8.1 for a diaphragm of 1 mm diameter.
CONCLUSION: The HRT is capable of measuring volumes with a reasonable accuracy. The accuracy does not depend on the width of the pupil over a wide range of pupil diameters. In cases of very small or very much dilated pupils the accuracy and reproducibility declines.

Mesh:

Year:  1994        PMID: 7967413     DOI: 10.1055/s-2008-1045499

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  1 in total

1.  Comparison of long-term fluctuations: laser scanning tomography versus automated perimetry.

Authors:  Jens Funk; Heiko Mueller
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-08-29       Impact factor: 3.117

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.