Literature DB >> 7952410

On-line computerized vectorcardiography monitoring of myocardial ischemia during coronary angioplasty: comparison with 12-lead electrocardiography.

S M Jensen1, G Johansson, G Osterman, S Reiz, U Näslund.   

Abstract

BACKGROUND: With new interventions minimizing ischemic myocardial injury, accurate and reliable techniques for the detection and continuous monitoring of myocardial ischemia are essential. We compared two techniques used for the detection of myocardial ischemia during coronary angioplasty: on-line computerized vectorcardiographic (cVCG) monitoring and the standard electrocardiography (ECG) leads or the complete 12-lead ECG.
METHODS: Thirty patients scheduled for routine angioplasty were included in the study. cVCG was recorded continuously. The electrodes were placed according to the lead system described by Frank and connected to a computerized system for on-line vectorcardiography. A 12-lead ECG was recorded simultaneously. The absolute variable spatial ST vector magnitude (ST-VM) and the relative variable spatial ST change vector magnitude (STC-VM) were calculated and compared with the standard 12-lead ECG for the detection of ischemia.
RESULTS: The sum of deviation in ST segment in all 12 standard ECG leads correlated closely with STC-VM, irrespective of which artery was occluded. STC-VM indicated ischemia during the first balloon inflation in 87% of the patients and demonstrated ischemia in more patients than the standard 12-lead ECG. Myocardial ischemia was not demonstrated by ST-VM in five out of 26 patients with ischemia according to STC-VM. In these cases, mainly directional vector changes and fewer changes in magnitude were observed.
CONCLUSION: Compared with 12-lead ECG, on-line cVCG is a more sensitive method of detecting myocardial ischemia during coronary angioplasty and the reading is easier and faster. Our results support STC-VM > or = 0.050 mV as the criterion for ischemia during angioplasty; ST-VM should be applied together with STC-VM.

Entities:  

Mesh:

Year:  1994        PMID: 7952410

Source DB:  PubMed          Journal:  Coron Artery Dis        ISSN: 0954-6928            Impact factor:   1.439


  3 in total

1.  Computerized vectorcardiography telemetry: a new device for continuous multilead ST-segment monitoring of ambulatory patients. A preliminary report.

Authors:  Bjarne Linde Nørgaard; Claus Sørensen; Thomas Larsen; Kristian Thygesen; Mikael Dellborg
Journal:  Ann Noninvasive Electrocardiol       Date:  2002-07       Impact factor: 1.468

2.  The influence of acute-phase levels of haemostatic factors on reperfusion and mortality in patients with acute myocardial infarction treated with streptokinase.

Authors:  Johan B Nilsson; Kurt Boman; Jan-Håkan Jansson; Torbjörn Nilsson; Ulf Näslund
Journal:  J Thromb Thrombolysis       Date:  2007-08-25       Impact factor: 2.300

3.  Ventricular depolarisation vectors in exercise induced myocardial ischaemia.

Authors:  Cameruddin W Vellani; Mohammad Yusuf; Sadia Mahmud; Satwat Hashmi
Journal:  Sci Rep       Date:  2017-11-07       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.