High-performance liquid chromatography and capillary electrophoresis of L- and D-carnitine by precolumn diastereomeric derivatization.

P De Witt1, R Deias, S Muck, B Galletti, D Meloni, P Celletti, A Marzo.   

Abstract

A new method for the simultaneous assay of D- and L-enantiomers of carnitine is described. The method is based on precolumn derivatization with (+)-1-(9-fluorenyl)ethyl chloroformate [(+)FLEC] producing a diastereomeric derivative which can be detected both by UV absorbance and fluorescence detection. Also acyl esters of carnitine can be processed with this method, after alkaline hydrolysis. The D-enantiomer of carnitine and acylcarnitine can be detected at a concentration as low as 0.2% in the raw material and in pharmaceuticals. Assays can be carried out using an autoinjector either by HPLC or capillary electrophoresis (CE) because the derivative proved to be very stable. Its application is proposed for the routine assay of the enantiomeric excess of L-carnitine and their acyl esters in pharmaceutical products.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7952085     DOI: 10.1016/0378-4347(94)80071-5

Source DB:  PubMed          Journal:  J Chromatogr B Biomed Appl        ISSN: 1572-6495


  2 in total

1.  Enantioselective micellar electrokinetic chromatography of dl-amino acids using (+)-1-(9-fluorenyl)-ethyl chloroformate derivatization and UV-induced fluorescence detection.

Authors:  Amir Prior; Erik van de Nieuwenhuijzen; Gerhardus J de Jong; Govert W Somsen
Journal:  J Sep Sci       Date:  2018-05-22       Impact factor: 3.645

2.  Validation of a Stability-Indicating RP-HPLC Method for Determination of l-Carnitine in Tablets.

Authors:  Roghaieh Khoshkam; Minoo Afshar
Journal:  Int Sch Res Notices       Date:  2014-10-22