Literature DB >> 7939387

Clinical findings on human blood with the KONE ISE for Mg2+.

M Maj-Zurawska1.   

Abstract

Ionized magnesium was measured using a clinical analyzer MICROLYTE 6 (KONE, Finland). Total magnesium was measured colorimetrically. The influence of anion-ligands on Mg-ISE responses was investigated. Ionized magnesium (iMg) concentration changes with pH were made to observe the accuracy of results. The dependence of iMg and iCa on heparin (used as anticoagulant in plasma samples) was made. Less than 40 units per 1 ml of blood decreased the concentration of iMg and iCa less than 1.5%. The range of iMg and total Mg in serum and plasma of healthy and ill adults (short bowel syndrome, myocardial infarct) was investigated and compared to each other. In both illnesses, the concentration of ionized and/or total magnesium is significantly different from the range for healthy people. The range of iMg and total Mg as well as the percentage of iMg against total Mg seem to be specific for given illness. The range of iMg and total Mg in blood serum of healthy children was also investigated.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7939387

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest Suppl        ISSN: 0085-591X


  2 in total

1.  Liquid heparin anticoagulant produces more negative bias in the determination of ionized magnesium than ionized calcium.

Authors:  Cheung Soo Shin; Chul Ho Chang; Jeong-Ho Kim
Journal:  Yonsei Med J       Date:  2006-04-30       Impact factor: 2.759

2.  Magnesium basics.

Authors:  Wilhelm Jahnen-Dechent; Markus Ketteler
Journal:  Clin Kidney J       Date:  2012-02
  2 in total