Automated and manual serum free thyroxine assays evaluated with equilibrium dialysis.

K Liewendahl1, L Melamies, T Helenius, S L Karonen, E Liljeström, M Välimäki, T Weber.   

Abstract

Two fully automated serum FT4 assays (Ciba Corning: ACS-180; Abbott Diagnostics: IMx) and two manual/semiautomated assays (Wallac: Delfia; Diagnostic Products Corp.: Coat-A-Count) were compared with a FT4 reference method based on equilibrium dialysis (ED) using routine clinical samples (n = 105-150). For the full range of FT4 concentrations the correlation with ED was good (r = 0.932-0.959), except for Coat-A-Count (r = 0.852), a single-step analogue-type method. Analytical inaccuracy of the automated methods at low FT4 concentrations was revealed by the differential plot: the IMx assay overestimated concentrations < 10 pmol l-1 and the ACS-180 assay concentrations < 6 pmol l-1. The between-assay CVs for the automated methods were not better than for the non-automated assays indicating problems of calibration curve stability for the automated assays.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7939380     DOI: 10.3109/00365519409087533

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  1 in total

1.  Assessment of levothyroxine sodium bioavailability: recommendations for an improved methodology based on the pooled analysis of eight identically designed trials with 396 drug exposures.

Authors:  Ingeborg Walter-Sack; Christof Clanget; Reinhard Ding; Christoph Goeggelmann; Vera Hinke; Matthias Lang; Johannes Pfeilschifter; Yorki Tayrouz; Karl Wegscheider
Journal:  Clin Pharmacokinet       Date:  2004       Impact factor: 6.447

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.