Literature DB >> 7932453

Pulse oximetry in the diagnosis of non-critical peripheral vascular insufficiency.

N F Couse1, C P Delaney, P G Horgan, J O'Keeffe, W P Joyce, T F Gorey, J M Fitzpatrick.   

Abstract

Pulse oximetry was used to detect return of pulsatile flow in 27 subjects during reactive hyperaemia following 3 min of total limb ischaemia induced by above knee tourniquet occlusion. Fourteen patients with exercise induced leg pain had 18 symptomatic limbs tested. Thirteen controls had 25 limbs tested. Return of pulsatile flow during reactive hyperaemia occurred within 20 s of tourniquet release in the 25 control limbs which was then regarded as normal. The mean time for return of pulsatile flow in 18 symptomatic limbs was 53 +/- 37 s (P < 0.05 versus controls). Three limbs had a normal value, two of which did not have peripheral vascular disease. Pulse oximetry correctly identified all 25 asymptomatic limbs and 15 of 16 patients with claudication secondary to peripheral vascular disease (PVD). This modification of the reactive hyperaemia test using the pulse oximeter is simple and quick to perform. It has potential as a non-invasive screening test for PVD, suitable for outpatient assessment.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7932453      PMCID: PMC1294763          DOI: 10.1177/014107689408700907

Source DB:  PubMed          Journal:  J R Soc Med        ISSN: 0141-0768            Impact factor:   18.000


  8 in total

1.  Doppler ankle pressure and reactive hyperemia in the diagnosis of arterial insufficiency.

Authors:  W C Johnson
Journal:  J Surg Res       Date:  1975-02       Impact factor: 2.192

2.  Postocclusion and postexercise flow velocity and ankle pressures in normals and marathon runners.

Authors:  F Mahler; L Koen; K H Johansen; E F Bernstein; A Fronek
Journal:  Angiology       Date:  1976-12       Impact factor: 3.619

3.  Blood flow limits and pulse oximeter signal detection.

Authors:  D Lawson; I Norley; G Korbon; R Loeb; J Ellis
Journal:  Anesthesiology       Date:  1987-10       Impact factor: 7.892

4.  Use of pulse oximetry for assessment of collateral arterial flow.

Authors:  G S Nowak; S S Moorthy; W L McNiece
Journal:  Anesthesiology       Date:  1986-04       Impact factor: 7.892

5.  Pulse oximetry: a new non-invasive assessment of peripheral arterial occlusive disease.

Authors:  W P Joyce; K Walsh; D B Gough; T F Gorey; J M Fitzpatrick
Journal:  Br J Surg       Date:  1990-10       Impact factor: 6.939

6.  Current status of noninvasive tests in the diagnosis of peripheral arterial disease.

Authors:  E F Bernstein; A Fronek
Journal:  Surg Clin North Am       Date:  1982-06       Impact factor: 2.741

7.  The pulse-reappearance time: an index of over-all blood flow impairment in the ischemic extremity.

Authors:  A Fronek; M Coel; E F Bernstein
Journal:  Surgery       Date:  1977-04       Impact factor: 3.982

8.  Poststress Doppler ankle pressures: a comparison of treadmill exercise with two other methods of induced hyperemia.

Authors:  J D Baker
Journal:  Arch Surg       Date:  1978-10
  8 in total