The influence of tape type and of skin preparation on the force required to dislodge angiocatheters.

N Patel1, C E Smith, A C Pinchak, D E Hancock.   

Abstract

The study evaluated the effects of different techniques used to secure intravenous (i.v.) catheters. An angiocatheter attached to standard i.v. tubing was taped to human forearm using a standard taping method. A calibrated piezoelectric force transducer was attached to the i.v. tubing. The force applied along the longitudinal axis to pull out the taped catheter was measured and recorded on paper. Three tape types, Curity, Leukopor and Transpore, were evaluated alone, with benzoin skin pretreatment and with mastisol pretreatment. A randomized 3 x 3 block design with 20 replications per block was utilized, and a total of 180 pullout tests were performed on two adult volunteers. Without pretreatment, the forces required to dislodge catheters were (means +/- SEM) 46 +/- 2, 37 +/- 2 and 38 +/- 2 Newtons for Curity, Leukopor and Transpore tape, respectively. Corresponding values for mastisol pretreatment (64 +/- 1, 64 +/- 3 and 52 +/- 3 Newtons) were greater (P < 0.05) for each tape compared with benzoin (54 +/- 3, 53 +/- 2 and 40 +/- 2 Newtons) and no pretreatment. The most frequent failure mode for Transpore tape was by tape fracture, for Curity tape was by separation from the skin of tape and catheter as a single unit, and for Leukopor tape was by catheter separation while tape remained attached to skin (P < 0.001). The data suggest that the application of mastisol prior to taping i.v. catheters with Curity or Leukopor tape helps to minimize the risk of accidental dislodgement.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7923524     DOI: 10.1007/BF03015631

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  3 in total

1.  The postoperative use of wound adhesives. Gum mastic versus benzoin, USP.

Authors:  C B Lesesne
Journal:  J Dermatol Surg Oncol       Date:  1992-11

2.  An evaluation of wound closure tapes.

Authors:  R L Moy; M B Quan
Journal:  J Dermatol Surg Oncol       Date:  1990-08

3.  Reinforcement of surgical adhesive strips.

Authors:  G R Mikhail; L Selak; S Salo
Journal:  J Dermatol Surg Oncol       Date:  1986-09
  3 in total
  1 in total

1.  Evaluation of different methods of securing intravenous catheters: measurement of forces during simulated accidental pullout.

Authors:  N Patel; C E Smith; A C Pinchak; D E Hancock
Journal:  Can J Anaesth       Date:  1995-06       Impact factor: 5.063

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.