Implications of a biphasic two-compartment model of constant flow ventilation for the clinical setting.

J R Hotchkiss1, P S Crooke, A B Adams, J J Marini.   

Abstract

PURPOSE: To investigate the theoretical effects of changing frequency (f), duty cycle (D), or end-inspiratory pause length on the distribution of ventilation and compartmental pressure in a heterogeneous, two compartment pulmonary model inflated by constant flow.
METHODS: Differential equations governing compartmental volume changes were derived and solved. Validation was conducted in a mechanical lung analogue with two mechanically independent compartments. Model predictions were then generated over wide ranges of f, D, or end-inspiratory pause.
RESULTS: Disparity of compartmental end-expiratory pressure was identified as the primary mechanism by which changes in f, D, or pause alter the distribution of ventilation. Distribution of peak pressures was less sensitive to such changes. Compartmental ventilation was much less uniform than compartmental peak pressure. Ventilation could not be made entirely uniform by changes of f, D, or pause within the usual clinical range.
CONCLUSIONS: In a linear, two compartment model of the respiratory system, disparity of compartmental end-expiratory pressures is the primary mechanism by which changes of f, D, or pause alter the distribution of ventilation during inflation with constant flow. Ventilation is less evenly distributed than peak alveolar pressure, and there are limits to the beneficial effects on the distribution of ventilation to be gained from manipulations of machine settings.

Mesh:

Year:  1994        PMID: 7920978     DOI: 10.1016/0883-9441(94)90022-1

Source DB:  PubMed          Journal:  J Crit Care        ISSN: 0883-9441            Impact factor:   3.425


  2 in total

1.  The Language of Caring: Quantitating Medical Practice Patterns using Symbolic Dynamics.

Authors:  J Paladino; A M Kaynar; P S Crooke; J R Hotchkiss
Journal:  Math Model Nat Phenom       Date:  2010-04-28       Impact factor: 4.157

2.  Optimal determination of respiratory airflow patterns using a nonlinear multicompartment model for a lung mechanics system.

Authors:  Hancao Li; Wassim M Haddad
Journal:  Comput Math Methods Med       Date:  2012-06-08       Impact factor: 2.238

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.