Literature DB >> 7919596

Roller and centrifugal pumps compared in vitro with regard to haemolysis, granulocyte and complement activation.

O Moen1, E Fosse, J Bråten, C Andersson, M K Fagerhol, P Venge, K Høgåsen, T E Mollnes.   

Abstract

A Biomedicus centrifugal pump and a Polystan roller pump were compared in vitro with regard to differences in haemolysis, granulocyte and complement activation. Six circuits of tubing and oxygenators were connected to each pump. Heparinized fresh human blood was circulated for 72 hours in the systems. Blood samples were drawn at defined intervals. Haemolysis was assessed by determination of lactate dehydrogenase (LD) and potassium, and granulocyte activation by quantification of the granulocyte proteins calprotectin, lactoferrin and myeloperoxidase. Complement activation was assessed by measuring C3 activation products (C3b, iC3b and C3c), and the terminal C5b-9 complement complex (TCC). The results indicate more haemolysis and complement activation in the roller pump group, revealed by significantly higher concentrations of LD, potassium, C3 activation products and TCC. Calprotectin, lactoferrin and myeloperoxidase were all significantly increased in both groups, but the rise appeared earlier in the roller pump group. The concentrations of LD and potassium both correlated significantly with C3 activation products, indicating that complement activation may at least partly be responsible for the haemolysis.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7919596     DOI: 10.1177/026765919400900205

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  5 in total

1.  Vitamins Q and E, extracorporal circulation and hemolysis.

Authors:  J Karlsson; R Ronneberg; B Semb
Journal:  Mol Cell Biochem       Date:  1997-08       Impact factor: 3.396

Review 2.  Functional and clinical aspects of the myelomonocyte protein calprotectin.

Authors:  B Johne; M K Fagerhol; T Lyberg; H Prydz; P Brandtzaeg; C F Naess-Andresen; I Dale
Journal:  Mol Pathol       Date:  1997-06

Review 3.  The inflammatory response to extracorporeal membrane oxygenation (ECMO): a review of the pathophysiology.

Authors:  Jonathan E Millar; Jonathon P Fanning; Charles I McDonald; Daniel F McAuley; John F Fraser
Journal:  Crit Care       Date:  2016-11-28       Impact factor: 9.097

4.  Hydrogen gas with extracorporeal cardiopulmonary resuscitation improves survival after prolonged cardiac arrest in rats.

Authors:  Tai Yin; Lance B Becker; Rishabh C Choudhary; Ryosuke Takegawa; Muhammad Shoaib; Koichiro Shinozaki; Yusuke Endo; Koichiro Homma; Daniel M Rolston; Shuhei Eguchi; Tadashi Ariyoshi; Asami Matsumoto; Kentaro Oka; Motomichi Takahashi; Tomoaki Aoki; Santiago J Miyara; Mitsuaki Nishikimi; Junichi Sasaki; Junhwan Kim; Ernesto P Molmenti; Kei Hayashida
Journal:  J Transl Med       Date:  2021-11-16       Impact factor: 5.531

5.  Utility of D-dimers in COVID-19 Patients Requiring Extracorporeal Membrane Oxygenation.

Authors:  Aqsa Shakoor; Stacey Chen; Jonathan Hyde; Brendan Wu; Bridget Toy; Stephanie Chang; Zachary Kon; Greta Piper; Deane Smith
Journal:  ASAIO J       Date:  2022-05-24       Impact factor: 3.826

  5 in total