Literature DB >> 7919249

[The in-vivo diffusion measurements of the liver, kidneys, spleen and m. erector with an echo-planar imaging system in normal subjects].

M F Müller1, P V Prasad, B Siewert, R R Edelman.   

Abstract

A diffusion sensitive pulse sequence using a stimulated echo (STEAM) type excitation and echo-planar (EPI) readout was developed on a 1.5 T echo-planar MR imaging system. Using this sequence the authors measured the apparent diffusion coefficients (ADC) of water in normal human liver, spleen, muscle and kidney in 12 volunteers. ADCs measured in normal volunteers were: liver, 1.39 +/- 0.16 x 10(-3) mm2/s, spleen, 0.95 +/- 0.15 x 10(-3) mm2/s, muscle, 1.99 +/- 0.16 x 10(-3) mm2/s, kidney, 3.54 +/- 0.47 x 10(-3) mm2/s. A large scatter of the ADCs in the kidneys was found in the different degrees of hydration status among the volunteers and also due to inter-subject variability. With further clinical experience, in vivo diffusion measurements of the abdominal organs may prove helpful in the identification and classification of abdominal disease by MRI.

Entities:  

Mesh:

Year:  1994        PMID: 7919249     DOI: 10.1055/s-2008-1032527

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010