Literature DB >> 7896605

Evaluation of a symmetrically disposed Pitot tube flowmeter for measuring gas flow during exercise.

J Porszasz1, T J Barstow, K Wasserman.   

Abstract

We evaluated the effect of airflow and gas composition on the linearity of measurement of airflow by a new disposable flowmeter. The flowmeter is based on the principle of differential pressure measurement across two symmetrically disposed Pitot tubes. Nonlinearities arising from the pressure-to-airflow relationship and sensitivity to changes in gas density were linearized with appropriate software and monitoring of the gas composition. With room air used as the respired gas, the measured tidal volume from a piston pump assembly was consistently within 1-2% of the target tidal volume for each of five flowmeters tested across physiological ranges of flow. Changing gas densities by varying concentrations of O2, CO2, and N2 led to errors in tidal volume measurement that ranged up to 6-8%. However, because the errors were predictable, they were corrected by software to within 0.6% of the target volume. Measurement of minute ventilation during exercise was within 1-2% of that determined from bag collections. We conclude that this type of flowmeter can accurately measure exercise minute ventilation and has advantages over some other flowmeters because of its ruggedness, reproducibility, and ease of sterilization or replacement compared with other flowmeters.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7896605     DOI: 10.1152/jappl.1994.77.6.2659

Source DB:  PubMed          Journal:  J Appl Physiol (1985)        ISSN: 0161-7567


  9 in total

1.  Heart rate variability during cycloergometric exercise or judo wrestling eliciting the same heart rate level.

Authors:  François Cottin; François Durbin; Yves Papelier
Journal:  Eur J Appl Physiol       Date:  2003-10-14       Impact factor: 3.078

Review 2.  Automated metabolic gas analysis systems: a review.

Authors:  D J Macfarlane
Journal:  Sports Med       Date:  2001       Impact factor: 11.136

3.  Lack of agreement between gas exchange variables measured by two metabolic systems.

Authors:  Djordje G Jakovljevic; David Nunan; Gay Donovan; Lynette D Hodges; Gavin R H Sandercock; David A Brodie
Journal:  J Sports Sci Med       Date:  2008-03-01       Impact factor: 2.988

Review 4.  Open-circuit respirometry: a historical review of portable gas analysis systems.

Authors:  Duncan J Macfarlane
Journal:  Eur J Appl Physiol       Date:  2017-10-17       Impact factor: 3.078

5.  Evaluation of a Pitot type spirometer in helium/oxygen mixtures.

Authors:  S Søndergaard; S Kárason; S Lundin; O Stenqvist
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502

6.  Comparative study of a novel portable indirect calorimeter to a reference breath-by-breath instrument and its use in telemedicine settings.

Authors:  S Jimena Mora; Mark Sprowls; Vishal V Tipparaju; Courtney M Wheatley-Guy; Doina Kulick; Bruce Johnson; Xian Xiaojun; Erica Forzani
Journal:  Clin Nutr ESPEN       Date:  2021-09-30

7.  Cardiopulmonary responses to exercise in patients with hypertrophic cardiomyopathy.

Authors:  S Jones; P M Elliott; S Sharma; W J McKenna; B J Whipp
Journal:  Heart       Date:  1998-07       Impact factor: 5.994

8.  Performance and Submaximal Adaptations to Additional Speed-Endurance Training Vs. Continuous Moderate-Intensity Aerobic Training in Male Endurance Athletes.

Authors:  Vincenzo Rago; Peter Krustrup; Magni Mohr
Journal:  J Hum Kinet       Date:  2022-09-08       Impact factor: 2.923

9.  Disadvantages of Automated Respiratory Gas Exchange Analyzers.

Authors:  Juan José Ramos-Álvarez; Irma Lorenzo-Capellá; Francisco Javier Calderón-Montero
Journal:  Front Physiol       Date:  2020-02-07       Impact factor: 4.566

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.