| Literature DB >> 7892189 |
F Wittlich1, K Kohno, G Mies, D G Norris, M Hoehn-Berlage.
Abstract
NMR bolus track measurements were correlated with autoradiographically determined regional cerebral blood flow (rCBF). The NMR method is based on bolus infusion of the contrast agent gadolinium diethylenetriaminepentaacetate and high-speed T*2-sensitive NMR imaging. The first pass of the contrast agent through the image plane causes a transient decrease of the signal intensity. This time course of the signal intensity is transformed into relative concentrations of the contrast agent in each pixel. The mean transit time and relative blood flow and volume are calculated from such indicator dilution curves. We investigated whether this NMR technique correctly expresses the relative rCBF. The relative blood flow data, calculated from NMR bolus track experiments, and the absolute values of iodo[14C]antipyrine autoradiography were compared. A linear relationship was observed, indicating the proportionality of the transient NMR signal change with CBF. Excellent interindividual reproducibility of calibration constants is observed (r = 0.963). For a given NMR protocol, bolus track measurements calibrated with autoradiography after the experiment allow determination of absolute values for rCBF and regional blood volume.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7892189 PMCID: PMC42379 DOI: 10.1073/pnas.92.6.1846
Source DB: PubMed Journal: Proc Natl Acad Sci U S A ISSN: 0027-8424 Impact factor: 11.205