The ambulatory blood pressure in normotensive and hypertensive subjects: results from an international database.

L Thijs1, J Staessen, E O'Brien, A Amery, N Atkins, P Baumgart, P De Cort, J P Degaute, P Dolenc, R De Gaudemaris.   

Abstract

OBJECTIVE: To delineate more precisely an operational threshold for making clinical decisions based on ambulatory blood pressure (ABP) measurement by studying the ABP in subjects who were diagnosed as either normotensive or hypertensive by conventional blood pressure (CBP) measurement.
SUBJECTS: Twenty-four research groups recruited 7069 subjects. Of these, 4577 were normotensive (systolic CBP < or = 140 mmHg and diastolic CBP < or = 90 mmHg) and 1773 were hypertensive (systolic CBP > or = 160 mmHg and/or diastolic CBP > or = 90 mmHg). Of the latter, 1324 had systolic and 1310 had diastolic hypertension.
RESULTS: Ninety-five percent of the normotensive subjects had a 24-h ABP below (systolic and diastolic, respectively) 133 and 82 mmHg. Of the patients with systolic hypertension, 24% had a 24-h systolic ABP of < 133 mmHg. Similarly, 30% of those with diastolic hypertension had a 24-h diastolic ABP of < 82 mmHg. The probability that hypertensive patients had a 24-h ABP below these thresholds was higher in women than in men, increased with age and was 2- to 4-fold greater if the CBP of the patient had been measured at only one visit and if fewer than 3 CBP measurements had been averaged to establish the diagnosis of hypertension. By contrast, for each 10-mmHg increment in systolic CBP, this probability decreased by 54% for the 24-h systolic ABP and by 25% for the 24-h diastolic ABP, and for each 5 mmHg increment in diastolic CBP it increased by 6 and 9%, respectively.
CONCLUSION: The ABP distributions of the normotensive subjects included in the present international database were not materially different from those in previous reports in the literature. One-fifth to more than one-third of the hypertensive patients had an ABP which was below the 95th centile of the ABP in normotensive subjects, but this proportion decreased if the hypertensive patients had shown a higher CBP upon repeated measurement. The prognostic implications of elevated CBP in the presence of normal ABP remain to be determined.

Entities:  

Mesh:

Year:  1995        PMID: 7885522     DOI: 10.1016/0300-2977(94)00057-g

Source DB:  PubMed          Journal:  Neth J Med        ISSN: 0300-2977            Impact factor:   1.422


  3 in total

1.  Pharmacokinetic and pharmacodynamic modelling of arterial haemodynamic effects of terazosin in healthy volunteers.

Authors:  Miguel Angel Campanero; Belén Sádaba; Maria José Muñoz-Juarez; Emilio García Quetglas; Jose Ramón Azanza
Journal:  Clin Drug Investig       Date:  2008       Impact factor: 2.859

2.  Ambulatory blood pressure monitoring for hypertension in general practice.

Authors:  R S Taylor; J Stockman; D Kernick; D Reinhold; A C Shore; J E Tooke
Journal:  J R Soc Med       Date:  1998-06       Impact factor: 5.344

Review 3.  Ambulatory blood pressure monitoring over 24 h: A Latin American Society of Hypertension position paper-accessibility, clinical use and cost effectiveness of ABPM in Latin America in year 2020.

Authors:  Ramiro A Sánchez; José Boggia; Ernesto Peñaherrera; Weimar Sebba Barroso; Eduardo Barbosa; Raúl Villar; Leonardo Cobos; Rafael Hernández Hernández; Jesús Lopez; José Andrés Octavio; José Z Parra Carrillo; Agustín J Ramírez; Gianfranco Parati
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-02-12       Impact factor: 3.738

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.