Literature DB >> 7877566

On the durability of pyrolytic carbon in vivo.

A D Haubold1.   

Abstract

The durability of pyrolytic carbon heart valve components was examined from the point of view of number implanted, documentation of wear on explanted components, and from the aspect of fatigue. Failures of pyrolytic carbon components were found to be few in number. A model describing the time course of events and successful usage of pyrolytic carbon components in heart valves was developed. The model is based on the yearly shipments of pyrolytic carbon components or valves, starting with 1969. The model indicates that about 2 million components have been successfully implanted, resulting in accumulative experience of over 10 million patient years. The wear of pyrolytic carbon, based on analysis of explanted heart valves, was found to be minimal. Additional wear data obtained on explanted components confirms the earlier observation that wear in vivo in less than that observed in vitro. The recently discovered fatigue behavior of pyrolytic carbon was found to have no demonstrated practical impact on the durability of pyrolytic carbon components used in existing mechanical heart valves.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7877566

Source DB:  PubMed          Journal:  Med Prog Technol        ISSN: 0047-6552


  7 in total

Review 1.  Advances in Proximal Interphalangeal Joint Arthroplasty: Biomechanics and Biomaterials.

Authors:  Andy F Zhu; Paymon Rahgozar; Kevin C Chung
Journal:  Hand Clin       Date:  2018-05       Impact factor: 1.907

2.  Pyrocarbon finger joint implant: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2004-03-01

3.  Intraoperative fracture of a pyrocarbon PIP total joint-a case report.

Authors:  Martin Skie; Nicholas Gove; Despina Ciocanel
Journal:  Hand (N Y)       Date:  2007-03-27

4.  Outcomes of pyrolytic carbon arthroplasty for the proximal interphalangeal joint.

Authors:  Kevin C Chung; Ashwin N Ram; Melissa J Shauver
Journal:  Plast Reconstr Surg       Date:  2009-05       Impact factor: 4.730

5.  Biological and functional evaluation of a novel pyrolytic carbon implant for the treatment of focal osteochondral defects in the medial femoral condyle: assessment in a canine model.

Authors:  Samantha L Salkeld; Laura P Patron; Joan C Lien; Stephen D Cook; Deryk G Jones
Journal:  J Orthop Surg Res       Date:  2016-12-01       Impact factor: 2.359

6.  Tendon versus Pyrocarbon Interpositional Arthroplasty in the Treatment of Trapeziometacarpal Osteoarthritis.

Authors:  Won-Taek Oh; Yong-Min Chun; Il-Hyun Koh; Jong-Kwan Shin; Yun-Rak Choi; Ho-Jung Kang
Journal:  Biomed Res Int       Date:  2019-07-22       Impact factor: 3.411

7.  Pyrocardan® interpositional arthroplasty for trapeziometacarpal osteoarthritis: a minimum four year follow-up.

Authors:  Francesco Smeraglia; Morena Anna Basso; Giulia Famiglietti; Andrea Cozzolino; Giovanni Balato; Alessio Bernasconi
Journal:  Int Orthop       Date:  2022-06-08       Impact factor: 3.479

  7 in total