A computerized control system for cardiopulmonary bypass.

T Beppu1, Y Imai, Y Fukui.   

Abstract

A pump control system using a microcomputer for cardiopulmonary bypass has been developed. The parameters monitored by the computer are central venous pressure, aortic pressure, blood volume in a reservoir, and collapsing of a small vinyl bag in a venous withdrawal tube. Both pumps in the arterial infusion and the venous withdrawal lines are automatically controlled through an interface unit throughout cardiopulmonary bypass. The system regulates central venous pressure with a proportional-integral control algorithm to maintain intravascular volume constant. A screening algorithm is devised to remove artifacts introduced to the pressure measurement. During the weaning period, a perfusionist can select either the central venous pressure control or a reservoir blood volume control. Computerized pump perfusion was applied on 15 children during cardiac operations. Perfusion flow and central venous pressure were controlled satisfactorily with stable operation. Compared with conventional manually controlled pump perfusion, no significant difference was noted in mean aortic pressure, central venous pressure, net fluid balance, total urination, blood chemistry, and urinalysis. This system is useful and is expected to improve the safety of pump perfusion.

Entities:  

Mesh:

Year:  1995        PMID: 7877303     DOI: 10.1016/S0022-5223(95)70273-3

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  1 in total

1.  Robust and self-tuning blood flow control during extracorporeal circulation in the presence of system parameter uncertainties.

Authors:  B J E Misgeld; J Werner; M Hexamer
Journal:  Med Biol Eng Comput       Date:  2005-09       Impact factor: 2.602

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.