| Literature DB >> 7863560 |
R G Aarnink1, R J Giesen, A L Huynen, J J de la Rosette, F M Debruyne, H Wijkstra.
Abstract
This paper describes a practical method for automated determination of the contour of the prostate in ultrasonographic images. In this method, we use specific edge detection techniques, based on nonlinear Laplace filtering. Possible edges are located at zero-crossings of the second derivative of the image. The strength of the edge is reflected by the value of the gradient of the image at that location. Combining the information about location and strength, an edge intensity image is constructed from the initial ultrasonographic image. In our method, edge enhancement techniques are performed on the edge intensity image. Edges that actually represent a boundary are selected and linked: interpolation techniques are used to fill the gaps between detected boundary edges. The method for contour determination in ultrasonographic images is used for accurate volume measurements in an everyday clinical environment. Its computer implementation is fast, accurate (mean difference within 6% of exact volume) and easy to use.Mesh:
Year: 1994 PMID: 7863560 DOI: 10.1016/0301-5629(94)90028-0
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998