Comparisons between Hologic, Lunar and Norland dual-energy X-ray absorptiometers and other techniques used for whole-body soft tissue measurements.

P Tothill1, A Avenell, J Love, D M Reid.   

Abstract

OBJECTIVES: Assessment of the precision and accuracy of dual-energy X-ray absorptiometers (DXA) from three manufacturers, used for measuring soft-tissue composition, and comparability with each other and other techniques.
DESIGN: Measurements of an anthropomorphic model of variable composition and thickness. 11 volunteers measured with each instrument and by underwater weighing (UWW) and three brands of bioelectric impedance analysis (BIA) apparatus.
RESULTS: New software, introduced by each manufacturer during the course of the investigation, led to changes in measured fat proportion. The precision of determination of fat proportion by DXA in the (small) model was 3-4% (coefficient of variation), with little difference between brands. In vivo precision was 2-3%. In the model, measurements of % fat differed from the nominal values, but variation with thickness was small. There were significant mean differences of total fat proportion in the volunteers between pairs of DXA instruments of 2.6-6.3% fat. The SDs of the differences were 1.8-2.9% fat. Regional differences were greater, with trunk % fat being particularly underestimated by Hologic relative to Lunar and Norland. Compared with UWW, mean % fat was the same for Hologic, but higher for Lunar and Norland. SDs of 4% demonstrated inadequate agreement. The differences varied with proportion of bone in lean tissue, questioning the assumption of constant density of lean tissue in UWW. There were no mean differences of % fat between the BIA instruments and DXA and UWW, but SDs of 3-6% suggest that BIA using these instruments does not offer an acceptable accuracy in estimating fat proportion.
CONCLUSIONS: UWW has limitations as a reference method. DXA is a useful technique, but its limitations, particularly regarding assumptions about fat distribution, must be borne in mind. The differences of fat proportion recorded by the three DXA instruments are such as to preclude interchangeability in measurements of individual subjects or in clinical trials.

Entities:  

Mesh:

Year:  1994        PMID: 7859696

Source DB:  PubMed          Journal:  Eur J Clin Nutr        ISSN: 0954-3007            Impact factor:   4.016


  17 in total

1.  Pediatric in vivo cross-calibration between the GE Lunar Prodigy and DPX-L bone densitometers.

Authors:  Nicola J Crabtree; N J Shaw; C M Boivin; B Oldroyd; J G Truscott
Journal:  Osteoporos Int       Date:  2005-10-19       Impact factor: 4.507

2.  Long-term GH treatment of GH-deficient adults: comparison between one and two daily injections.

Authors:  C Höybye; M Rudling
Journal:  J Endocrinol Invest       Date:  2006-12       Impact factor: 4.256

3.  Estimating the Basal metabolic rate from fat free mass in individuals with motor complete spinal cord injury.

Authors:  S M Chun; H-R Kim; H I Shin
Journal:  Spinal Cord       Date:  2017-05-23       Impact factor: 2.772

4.  Familial resemblance of body composition, physical activity, and resting metabolic rate in pre-school children.

Authors:  Kurosh Djafarian; John R Speakman; Joanne Stewart; Diane M Jackson
Journal:  Rep Biochem Mol Biol       Date:  2013-10

5.  Validation of a body condition scoring system in rhesus macaques (Macaca mulatta): assessment of body composition by using dual-energy X-ray absorptiometry.

Authors:  Laura Summers; Karen J Clingerman; Xiaowei Yang
Journal:  J Am Assoc Lab Anim Sci       Date:  2012-01       Impact factor: 1.232

6.  Predicting skeletal muscle mass from dual-energy X-ray absorptiometry in Japanese prepubertal children.

Authors:  T Midorikawa; M Ohta; Y Hikihara; S Torii; S Sakamoto
Journal:  Eur J Clin Nutr       Date:  2017-04-05       Impact factor: 4.016

7.  Validation of fan beam dual energy x ray absorptiometry for body composition assessment in adults aged 18-45 years.

Authors:  J Norcross; M D Van Loan
Journal:  Br J Sports Med       Date:  2004-08       Impact factor: 13.800

8.  Reference data and percentile curves of body composition measured with dual energy X-ray absorptiometry in healthy Chinese children and adolescents.

Authors:  Bin Guo; Yi Xu; Jian Gong; Yongjin Tang; Jingjie Shang; Hao Xu
Journal:  J Bone Miner Metab       Date:  2014-10-16       Impact factor: 2.626

9.  Relationship between body composition and bone mineral density in women with and without osteoporosis: relative contribution of lean and fat mass.

Authors:  Saverio Gnudi; Emanuela Sitta; Nicoletta Fiumi
Journal:  J Bone Miner Metab       Date:  2007-08-25       Impact factor: 2.626

10.  Age trends of bone mineral density and percentile curves in healthy Chinese children and adolescents.

Authors:  Bin Guo; Yi Xu; Jian Gong; Yongjin Tang; Hao Xu
Journal:  J Bone Miner Metab       Date:  2013-01-30       Impact factor: 2.626

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.