Literature DB >> 7839843

Increased metal release from cemented femoral components made of titanium alloy. 19 hip prostheses followed with radiostereometry (RSA).

J Kärrholm1, W Frech, K G Nilsson, F Snorrason.   

Abstract

In 19 patients the concentrations of metal were measured in serum, urine and joint fluid 2 years after implantation of uncemented commercially pure titanium acetabular cups and cemented or uncemented femoral components made of titanium alloy. A ceramic against the polyethylene articulation was used. The fixation of the components was followed with radiostereometry (RSA). Samples from 12 patients scheduled for hip or knee prostheses and without any metallic implant were used as controls. High levels of titanium were found in cemented hips and when large acetabular cups had been inserted. Increased aluminum levels were also noted in the cemented hips. Vanadium was not detected in any of the samples. Micromotions were detected in most of the implants, but the magnitude of these movements could not be used to predict the release of metal into the synovial fluid.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7839843     DOI: 10.3109/17453679408994612

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  4 in total

1.  Causes of titanium release from plate and screws implanted in rabbits.

Authors:  Y Mu; T Kobayashi; K Tsuji; M Sumita; T Hanawa
Journal:  J Mater Sci Mater Med       Date:  2002-06       Impact factor: 3.896

2.  Ten-year outcome of serum metal ion levels after primary total hip arthroplasty: a concise follow-up of a previous report*.

Authors:  Brett R Levine; Andrew R Hsu; Anastasia K Skipor; Nadim J Hallab; Wayne G Paprosky; Jorge O Galante; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2013-03-20       Impact factor: 5.284

3.  A review of the biologic effects of spine implant debris: Fact from fiction.

Authors:  Nadim James Hallab
Journal:  SAS J       Date:  2009-12-01

4.  A new automated way to measure polyethylene wear in THA using a high resolution CT scanner: method and analysis.

Authors:  Gerald Q Maguire; Marilyn E Noz; Henrik Olivecrona; Michael P Zeleznik; Lars Weidenhielm
Journal:  ScientificWorldJournal       Date:  2014-01-22
  4 in total