Early changes in defibrillation threshold following implantation of a nonthoracotomy system in dogs.

B D Halperin1, D W Haupt, J H Lemmer, S R Holcomb, R P Oliver, M J Silka.   

Abstract

BACKGROUND: Nonthoracotomy systems are rapidly becoming the preferred surgical method for implantation of cardioverter defibrillators. Testing is performed at the time of implantation to insure an adequate margin of safety for defibrillation. However, this safety margin may change with lead maturation. This study evaluated changes in defibrillation threshold following implantation of a nonthoracotomy system. METHODS AND
RESULTS: Ten dogs underwent implantation of a nonthoracotomy system consisting of a single catheter with a distal coil electrode in the right ventricular apex and a proximal coil electrode in the superior vena cava forming a common anode with a subcutaneous patch over the left thorax. Defibrillation threshold testing, using a biphasic waveform, was performed on each animal under general anesthesia at implantation (day 1) and subsequently on postoperative days 3, 7, 10, 17, 24, 31, 38, and 45. E50, the energy associated with a 50% likelihood of successful defibrillation, was determined at each setting. The mean E50 was 12.2 +/- 1.1 J at the time of implantation, increasing 36% to 16.8 +/- 2.0 J by day 38 (P < 0.01). Individual increases in E50 of 10-12 J were observed in four animals.
CONCLUSIONS: Energy requirements for defibrillation with a nonthoracotomy system increase during the early postoperative period, with the highest defibrillation threshold observed at 38 days. This increase may be applicable to humans and should be considered when selecting an adequate energy safety margin for defibrillation at time of implantation.

Entities:  

Mesh:

Year:  1994        PMID: 7838785     DOI: 10.1111/j.1540-8159.1994.tb03744.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Pacing threshold increase in nonthoracotomy implantable defibrillator leads: implications for battery longevity and margin of safety.

Authors:  A E Epstein; V J Plumb; K A Kirk; G N Kay
Journal:  J Interv Card Electrophysiol       Date:  1997-09       Impact factor: 1.900

2.  [Influence of waveform and configuration of electrodes on the defibrillation threshold of implantable cardioverter-defibrillators].

Authors:  M Block; D Hammel; G Breithardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.