Experience with two different nonthoracotomy systems for implantable defibrillator in 170 patients.

J S Sra1, A Natale, K Axtell, C Maglio, M Jazayeri, S Deshpande, A Dhala, Z Blanck, M Akhtar.   

Abstract

Implantation of a nonthoracotomy system (Medtronic PCD or CPI Endotak) was attempted in 170 patients with ventricular tachycardia (VT) or ventricular fibrillation (VF) not requiring concomitant cardiac surgery. A nonthoracotomy system could be successfully implanted in 95 of the 115 patients with the PCD system and 49 of 55 patients receiving the Endotak lead system. In 26 patients with failed nonthoracotomy system because of defibrillation threshold (DFT) > 25 joules (J), an epicardial system was implanted at the same setting. Patients receiving the two lead systems were comparable with regard to age, sex, and ejection fraction. However, since the PCD system offers tiered therapy multiprogrammable options, all attempts were made to implant this lead system in patients with VT that could be pace terminated. Mean DFT (15 +/- 4.7 vs 17 +/- 4.6 J; P = 0.03) and implant time (2.5 +/- 0.6 vs 3.3 +/- 0.7 hours; P = 0.02) were less with the Endotak lead system. There was no perioperative mortality. During a mean follow-up of 20 +/- 4 months, there were eight instances of lead dislodgment in patients receiving the PCD system. There were four nonsudden cardiac deaths and one sudden death in the Endotak group and three nonsudden deaths in the PCD group. Sudden cardiac death and total survival using the intention-to-treat analysis during this follow-up period were 99% and 95%, respectively. In conclusion, successful implantation, perioperative mortality, and survival rate are comparable with both lead systems; however, incorporating two defibrillating electrodes in one lead minimizes lead dislodgment and reduces implant time.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7838782     DOI: 10.1111/j.1540-8159.1994.tb03741.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

Review 1.  Is primary antiarrhythmic drug therapy for ventricular arrhythmias obsolete?

Authors:  R Cooley; M Akhtar; J Sra
Journal:  Curr Cardiol Rep       Date:  1999-11       Impact factor: 2.931

2.  Role of proximal electrode position in transvenous ventricular defibrillation.

Authors:  J Nitta; D S Khoury
Journal:  Ann Biomed Eng       Date:  1996 May-Jun       Impact factor: 3.934

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.