Abstract

Emergency situations, such as acute hydrocephalus or ventricular hemorrhage, require immediate and reliable treatment by ventriculostomy. The method used has to be standardized, applicable in a fast manner, and associated with only minimum risk of infection. We present a recently developed set for external ventriculostomy, which meets the above-standing requirements, and which consists of a screw with self-biting thread, a metal cannula, and a special screwdriver. Ventriculostomy can be performed easily within 2 minutes, and the system can be fixed rigidly to the skull of the patient for a period of up to several weeks. Exchange of the cannula is possible within 1 minute. The system has been used in 90 cases so far, with a rate of possible infection of 1.1%.

Entities:  

Mesh:

Year:  1994        PMID: 7838355     DOI: 10.1227/00006123-199411000-00030

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  1 in total

1.  MRI-compatible titanium ventriculostomy kit: technical note.

Authors:  C Schaller; C Pavlidis
Journal:  Acta Neurochir (Wien)       Date:  1996       Impact factor: 2.216