Literature DB >> 7835512

Bitewing-based alveolar bone densitometry: digital imaging resolution requirements.

C F Hildebolt1, T Q Bartlett, B S Brunsden, N L Hente, M J Gravier, R K Walkup, M K Shrout, M W Vannier.   

Abstract

The resolution requirements were determined for detection of incremental alveolar bone loss from digitized Ektaspeed radiographs. Ten clinical radiographs were examined with a calibrated optical microscope to measure the smallest feature of interest discernible for alveolar bone. Images of trabeculae > 100 microns in diameter could be identified, but smaller ones could not be resolved. The Nyquist Criterion requires sampling with 50 microns (or smaller) pixels to measure such features. Numerous 25 microns Ag aggregates were present. Fifty microns resolution is a practical compromise between noise level and feature resolution. In another example of 10 bitewings digitized at 8 bit grey scale, about half the available range was used for alveolar bone, resulting in uncertainty, over a range of 2 optical density (OD) units, of about 1.42% at the average OD for alveolar bone (1.1). With the same radiographs digitized at 12 bit resolution, over 2000 of 4096 grey scales were used with a corresponding uncertainty of about 0.09%. Twelve-bit precision was also able to resolve smaller increments in an aluminium stepwedge than was 8 bit precision. Twelve-bit grey scale and 50 microns spatial resolution are recommended for alveolar bone densitometry performed with Ektaspeed radiographs.

Entities:  

Mesh:

Year:  1994        PMID: 7835512     DOI: 10.1259/dmfr.23.3.7835512

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  2 in total

1.  Image resolution and exposure time of digital radiographs affects fractal dimension of periapical bone.

Authors:  B Güniz Baksi; Aleš Fidler
Journal:  Clin Oral Investig       Date:  2011-11-29       Impact factor: 3.573

2.  Measurement accuracy and perceived quality of imaging systems for the evaluation of periodontal structures.

Authors:  B Güniz Baksi
Journal:  Odontology       Date:  2008-07-27       Impact factor: 2.634

  2 in total