Literature DB >> 7825754

A study of optimal configuration and control of a multi-chamber balloon for intraaortic balloon pumping.

J Bai1, H Lin, Z Yang, X Zhou.   

Abstract

Balloon configuration and control scheme are important for the optimization of assistance of the failing heart with an intraaortic balloon pumping device. In this work, the configuration of a multi-chamber balloon and control schemes have been investigated by using a hemodynamic model and computer simulation methods. Following the simulation study, physical testing and animal experiments were performed to demonstrate the simulation results. Results show that the optimal configuration and controlled multi-chamber balloon can provide better assistance to the failing heart. Based on the simulation and experimental results, it was found that the shape of the rear chamber of a multi-chamber balloon is critical. The optimal control scheme was to inflate the rear chamber first and deflate it last.

Mesh:

Year:  1994        PMID: 7825754     DOI: 10.1007/bf02367088

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  9 in total

1.  Computer simulation of the mechanically-assisted failing canine circulation.

Authors:  O Barnea; T W Moore; D Jaron
Journal:  Ann Biomed Eng       Date:  1990       Impact factor: 3.934

2.  A closed-loop control scheme for intraaortic balloon pumping.

Authors:  J A Zelano; J K Li; W Welkowitz
Journal:  IEEE Trans Biomed Eng       Date:  1990-02       Impact factor: 4.538

3.  Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle.

Authors:  H Suga; K Sagawa
Journal:  Circ Res       Date:  1974-07       Impact factor: 17.367

4.  Clinical experience with a new cardiac assist device. The dual-chambered intra-aortic balloon assist.

Authors:  D Bregman; R H Goetz
Journal:  J Thorac Cardiovasc Surg       Date:  1971-10       Impact factor: 5.209

5.  The effect of synchronous unidirectional intra-aortic balloon pumping on hemodynamics and coronary blood flow in cardiogenic shock.

Authors:  D Bregman; D C Kripke; R H Goetz
Journal:  Trans Am Soc Artif Intern Organs       Date:  1970

6.  Left heart assist system: intra-aortic balloom pump.

Authors:  M J Buckley; J D Laird; P N Madras; R T Jones; A R Kantrowitz; W G Austen
Journal:  Surg Clin North Am       Date:  1969-06       Impact factor: 2.741

7.  Control of intraaortic balloon pumping: theory and guidelines for clinical applications.

Authors:  D Jaron; T W Moore; P He
Journal:  Ann Biomed Eng       Date:  1985       Impact factor: 3.934

8.  Evidence and quantitation of left ventricular systolic resistance.

Authors:  S G Shroff; J S Janicki; K T Weber
Journal:  Am J Physiol       Date:  1985-08

9.  Theoretical considerations regarding the optimization of cardiac assistance by intraaortic balloon pumping.

Authors:  D Jaron; T W Moore; P He
Journal:  IEEE Trans Biomed Eng       Date:  1983-03       Impact factor: 4.538

  9 in total
  1 in total

1.  Newly Shaped Intra-Aortic Balloons Improve the Performance of Counterpulsation at the Semirecumbent Position: An In Vitro Study.

Authors:  Christina Kolyva; John R Pepper; Ashraf W Khir
Journal:  Artif Organs       Date:  2016-08       Impact factor: 3.094

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.