| Literature DB >> 7825754 |
J Bai1, H Lin, Z Yang, X Zhou.
Abstract
Balloon configuration and control scheme are important for the optimization of assistance of the failing heart with an intraaortic balloon pumping device. In this work, the configuration of a multi-chamber balloon and control schemes have been investigated by using a hemodynamic model and computer simulation methods. Following the simulation study, physical testing and animal experiments were performed to demonstrate the simulation results. Results show that the optimal configuration and controlled multi-chamber balloon can provide better assistance to the failing heart. Based on the simulation and experimental results, it was found that the shape of the rear chamber of a multi-chamber balloon is critical. The optimal control scheme was to inflate the rear chamber first and deflate it last.Mesh:
Year: 1994 PMID: 7825754 DOI: 10.1007/bf02367088
Source DB: PubMed Journal: Ann Biomed Eng ISSN: 0090-6964 Impact factor: 3.934