Literature DB >> 7820832

Long-term follow-up of the Günther basket inferior vena cava filter: does mechanical instability cause complications?

C D Becker1, H M Hoogewoud, P Felder, I Gal, P A Ruijs, J Triller.   

Abstract

PURPOSE: The Günther basket inferior vena cava filter (GBF) has been withdrawn from the market because of its mechanical instability, but a number of patients still live with the device.
METHODS: In a two-center study, we evaluated the long-term follow-up of the GBF, based on clinical data in 78 patients, and on imaging studies including plain radiographs, Doppler ultrasound, angiography, or computed tomography (CT) in 50 patients.
RESULTS: In a mean period of 3 years, pulmonary embolism was diagnosed in five patients (6.4%), with an overall rate of 0.02 embolic episodes per patient per year. None of these patients required hospitalization, and there were no deaths due to pulmonary embolism. Inferior vena cava thrombosis was documented in three patients (3.9%), and occlusive venous thrombosis at the access site in seven patients (9%). Spontaneous migration was documented in 43% of the examined filters and spontaneous disruption in 77%. Dislocated filter fragments were localized by CT in the adjacent retroperitoneum in 11%, in the aortic lumen in 2%, and in a peripheral pulmonary artery in 7%. None of the patients had symptoms attributable to filter migration or disruption.
CONCLUSION: Our results indicate that the rate of clinically relevant complications with the GBF is no higher than with other vena cava filters. Because mechanical instability of the GBF had no clinical consequences, we conclude that patients who live with this device may be observed and treated in a manner similar to patients with other vena cava filters.

Entities:  

Mesh:

Year:  1994        PMID: 7820832     DOI: 10.1007/bf00192446

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  20 in total

1.  Cephalic migration of the bird's nest inferior vena caval filter: report of two cases.

Authors:  P A Rogoff; A D Hilgenberg; S L Miller; S M Stephan
Journal:  Radiology       Date:  1992-09       Impact factor: 11.105

2.  Spontaneous disruption of two Greenfield vena caval filters.

Authors:  W Lang; H Schweiger; R Fietkau; K Hofmann-Preiss
Journal:  Radiology       Date:  1990-02       Impact factor: 11.105

Review 3.  Percutaneous inferior vena caval filters.

Authors:  G S Dorfman
Journal:  Radiology       Date:  1990-03       Impact factor: 11.105

4.  [Strut fractures in the long-term follow-up of Günther cava filters--the results after 64 filter implantations].

Authors:  M Urban; W Hruby; W B Winkler; H Mosser; M Baldt
Journal:  Rofo       Date:  1992-04

5.  Caudal migration of the Günther vena caval filter.

Authors:  P A Schneider; L Parmeggiani; J C Piguet; H Bounameaux
Journal:  Radiology       Date:  1989-11       Impact factor: 11.105

6.  Percutaneous insertion of the Kimray-Greenfield filter: incidence of femoral vein thrombosis.

Authors:  A Kantor; S Glanz; D H Gordon; S J Sclafani
Journal:  AJR Am J Roentgenol       Date:  1987-11       Impact factor: 3.959

7.  Migration of the Simon nitinol vena cava filter to the chest.

Authors:  J S LaPlante; F M Contractor; P M Kiproff; M B Khoury
Journal:  AJR Am J Roentgenol       Date:  1993-02       Impact factor: 3.959

8.  Current indications for and results of Greenfield filter placement.

Authors:  L J Greenfield
Journal:  J Vasc Surg       Date:  1984-05       Impact factor: 4.268

9.  Access-site thrombosis after placement of inferior vena cava filters with 12-14-F delivery sheaths.

Authors:  C P Molgaard; E K Yucel; S C Geller; T A Knox; A C Waltman
Journal:  Radiology       Date:  1992-10       Impact factor: 11.105

10.  [Percutaneous transluminal implantation of the Günther cava filter: experience based on 22 patients].

Authors:  F Grabenwöger; W Dock; V Metz; K Eibenberger; H Magometschnigg
Journal:  Rofo       Date:  1991-11
View more
  1 in total

Review 1.  Inferior vena cava filters for recurrent thrombosis: current evidence.

Authors:  Salil H Patel; Rima Patel
Journal:  Tex Heart Inst J       Date:  2007
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.