Literature DB >> 7819385

Automated analytical systems for drug development studies. I--A system for the determination of drug stability.

K P Shah1, J Zhou, R Lee, R L Schowen, R Elsbernd, J M Ault, J F Stobaugh, M Slavik, C M Riley.   

Abstract

An automated system consisting of a pH-stat, microdialysis sampling and a liquid chromatograph was assembled to measure the rate of rapid chemical reactions. 2',3',5'-Triacetyl-6-azauridine was used as a model compound to validate the performance of the automated system. Buffer catalysis was minimized by using a non-catalytic concentration of borate buffer along with a pH-stat to maintain the pH during the kinetic run. The microdialysis sampling technique permitted sample quenching and buffering of the solutions to a pH compatible with the LC column materials. The combination of microdialysis sampling and rapid LC analysis allowed reactions with a half-life of approximately 1 min to be sampled every 30 s. The rates of hydrolysis of the drug, measured at different conditions of temperature (37-70 degrees C) and pH (9.0-10.5) using the automated system, compared well with the previously determined values.

Entities:  

Mesh:

Year:  1994        PMID: 7819385     DOI: 10.1016/0731-7085(94)e0011-o

Source DB:  PubMed          Journal:  J Pharm Biomed Anal        ISSN: 0731-7085            Impact factor:   3.935