Chemical noise limits mass spectrometric detection of chloramphenicol (CAP) with electron capture ionization at low resolution, and makes CAP identification at concentrations of 5 parts per billion (ppb) difficult. Increasing the resolution from 1000 to 3500, however, was sufficient to separate the analyte signals from the noise signals, and resulted in a 100 times higher analytical sensitivity. The introduction of sweep gas in the ion source decreased the scattering of the quantitative results on average by a factor of 7, and thereby improved the precision of the analyses to an acceptable level (CV < 10%). Under such conditions, CAP residues of 1.5 and 2.1 ppb in shrimp as determined by electron capture gas chromatography/mass spectrometry can readily be identified by monitoring four diagnostic ions.Entities: