| Literature DB >> 7802252 |
Abstract
A cyclic polyamine, 3-decyl-1,5,8-triazacyclodecane-2,4-dione (N3-cyclic amine), was used as the ionophore for a dibasic phosphate-selective electrode. This electrode exhibited a linear response between 1.0 mumol/L and 0.1 mol/L dibasic phosphate activity with a near-Nernstian slope of approximately -29 mV per activity decade. The electrode selectivity for dibasic phosphate over other commonly occurring anions was evaluated. A mechanism for the selectivity of the electrode toward HOP4(2-) ions is postulated to be a function of the size and charge of the N3-cyclic amine ionophore relative to the size and charge of HPO4(2-) ions. The electrode's superior selectivity and sensitivity make possible the direct measurement of phosphate activity in a wide variety of applications.Entities:
Mesh:
Substances:
Year: 1994 PMID: 7802252 DOI: 10.1021/ac00093a009
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986