Literature DB >> 7802056

Simple bladder filling with a cough stress test compared with subtracted cystometry for the diagnosis of urinary incontinence.

L L Wall1, A K Wiskind, P A Taylor.   

Abstract

OBJECTIVE: Our purpose was to compare the diagnostic efficacy of observing urine loss during simple bladder filling (without pressure measurement) and a cough stress test, with multichannel subtracted cystometry. STUDY
DESIGN: The urodynamic records of 77 women complaining of urinary incontinence were reviewed. All women had undergone a full evaluation that included a standardized history and physical examination, urinalysis and urine culture, uroflowmetry with measurement of postvoid residual urine, a cough stress test performed during and after simple retrograde bladder filling, and multichannel subtracted cystometry. The results of simple bladder filling and the cough stress test were recorded as "urge incontinence," "stress incontinence,"mixed incontinence," or "incontinence not demonstrated." The subtracted cystometrogram was then performed, and the urodynamic diagnoses were recorded as "detrusor instability," "genuine stress incontinence," "mixed incontinence," or "incontinence not demonstrated." The records were reviewed, and the results of simple bladder filling with a cough stress test were compared with those obtained by subtracted multichannel provocative cystometry.
RESULTS: With the subtracted multichannel cystometrogram used as the "gold standard" for diagnosis, the demonstration of "urge incontinence" during simple bladder filling had a sensitivity of 64% and a specificity of 86.8% for the demonstration of detrusor instability during cystometry, with a positive predictive value of 83.3% and a negative predictive value of 70.2%. The demonstration of "stress incontinence" during simple bladder filling had a sensitivity of 88.1% and a specificity of 77.1% for the demonstration of "genuine" stress incontinence during cystometry, with a positive predictive value of 82% and a negative predictive value of 84.4%.
CONCLUSIONS: The demonstration of urge incontinence during simple bladder filling is a reliable predictor of detrusor instability, but its absence is less reliable in excluding detrusor overactivity as a cause of urinary incontinence. Similarly, the clinical demonstration of stress incontinence during simple bladder filling is predictive of the presence of "genuine" stress incontinence during subtracted multichannel provocative cystometry. The inability to demonstrate stress incontinence during simple bladder filling is highly correlated with the absence of "genuine" stress incontinence during complex urodynamic testing. Simple bladder filling is a reliable method of diagnosing urinary incontinence. In many cases it can replace complex urodynamic testing, particularly if the proposed treatment for the condition carries a low level of risk, and is helpful in selecting patients who need more extensive evaluation.

Entities:  

Mesh:

Year:  1994        PMID: 7802056     DOI: 10.1016/0002-9378(94)90390-5

Source DB:  PubMed          Journal:  Am J Obstet Gynecol        ISSN: 0002-9378            Impact factor:   8.661


  9 in total

1.  Further characterization of the supine empty stress test for predicting low valsalva leak point pressures.

Authors:  Andrew J Walter; Jennifer A Thornton; Andrew C Steele
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2004-06-04

2.  Diagnostic accuracy of single channel cystometry for neurogenic bladder diagnosis following spinal cord injury: a pilot study.

Authors:  Akmal Hafizah Zamli; Kavitha Ratnalingam; Yusma Asni Yusmido; Kuo Ghee Ong
Journal:  Spinal Cord Ser Cases       Date:  2017-05-04

3.  Primary stress urinary incontinence: what to do and why.

Authors:  Jhansi Reddy; Marie Fidela R Paraiso
Journal:  Rev Obstet Gynecol       Date:  2010

4.  Reproducibility of a cough and jump stress test for the evaluation of urinary incontinence.

Authors:  G Horndalsveen Berild; S Kulseng-Hanssen
Journal:  Int Urogynecol J       Date:  2012-04-19       Impact factor: 2.894

5.  Retrospective study on tension-free vaginal tape obturator (TVT-O).

Authors:  Malathi Rajendra; How Chuan Han; Lih Charn Lee; Leng Aun Arthur Tseng; Heng Fok Wong
Journal:  Int Urogynecol J       Date:  2011-09-03       Impact factor: 2.894

6.  Long-term clinical outcomes with the retropubic tension-free vaginal tape (TVT) procedure compared to Burch colposuspension for correcting stress urinary incontinence (SUI).

Authors:  Bjørn Holdø; Margareta Verelst; Rune Svenningsen; Ian Milsom; Finn Egil Skjeldestad
Journal:  Int Urogynecol J       Date:  2017-04-24       Impact factor: 2.894

7.  How often does detrusor overactivity cause urinary leakage during a stress test in women with mixed urinary incontinence?

Authors:  Sigurd Kulseng-Hanssen; Kjartan Moe; Hjalmar A Schiøtz
Journal:  Int Urogynecol J       Date:  2013-02-02       Impact factor: 2.894

8.  Bladder Reconstruction with Human Amniotic Membrane in a Xenograft Rat Model: A Preclinical Study.

Authors:  Dimitri Barski; Holger Gerullis; Thorsten Ecke; Jin Yang; Gabriella Varga; Mihaly Boros; Isabel Pintelon; Jean-Pierre Timmermans; Thomas Otto
Journal:  Int J Med Sci       Date:  2017-03-11       Impact factor: 3.738

9.  Evaluation and outcome measures in the treatment of female urinary stress incontinence: International Urogynecological Association (IUGA) guidelines for research and clinical practice.

Authors:  G Ghoniem; E Stanford; K Kenton; C Achtari; R Goldberg; T Mascarenhas; M Parekh; K Tamussino; S Tosson; G Lose; E Petri
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-11-17
  9 in total