Literature DB >> 7795823

Simple method for the analysis of tenoxicam in human plasma using high-performance liquid chromatography.

J L Mason1, G J Hobbs.   

Abstract

A simple, rapid and cost-effective method for the determination of tenoxicam in human plasma is described, using ketorolac as the internal standard. The extraction procedure utilised 5% zinc sulphate and methanol. A nucleosil C18 column and 35:65 acetonitrile-water phosphate buffered mobile phase (pH 2.8) were used, with ultraviolet detection at 355 nm. The assay was linear in the range 40 ng/ml-10 micrograms/ml, with recovery of extraction ranging from 87 to 102%. The intra- and inter-assay reproducibility had coefficients of variation of 3.9-7.7 and 1.6% respectively. The limit of detection for this method was 40 ng/ml.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7795823     DOI: 10.1016/0378-4347(94)00534-c

Source DB:  PubMed          Journal:  J Chromatogr B Biomed Appl        ISSN: 1572-6495


  2 in total

1.  Improving tenoxicam solubility and bioavailability by cosolvent system.

Authors:  Ming-Kung Yeh; Li-Chien Chang; Andy Hong-Jey Chiou
Journal:  AAPS PharmSciTech       Date:  2009-02-18       Impact factor: 3.246

2.  Electroanalytical Determination of the Antiinflammatory Drug Tenoxicam in Pharmaceutical Dosage Forms.

Authors:  Fatma Ağin; Sena Atal
Journal:  Turk J Pharm Sci       Date:  2019-03-27
  2 in total