M R Feng1, P R Siersma, C A Strenkoski, D S Wright.
Abstract
A sensitive gradient high-performance liquid chromatographic (HPLC) method for the simultaneous quantitation of a dopamine autoreceptor agonist CI-1007 (I) and its metabolite PD 147693 (II) is described. Monkey plasma samples were purified by liquid-liquid extraction using hexane. Liquid chromatographic separation was achieved on two C18 analytical columns (installed in series) using gradient elution. Column effluent was monitored using a fluorescence detector programmed to change wavelengths at specified times. Minimum quantitation limits of I and II were 3.0 and 5.0 ng/ml, respectively, for a plasma sample volume of 0.100 ml. Linearity was demonstrated up to 300 ng/ml. The assay has been applied to the analysis of I and II in plasma from monkeys following intravenous and oral doses of I.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7795791 DOI: 10.1016/0378-4347(94)00506-z
Source DB: PubMed Journal: J Chromatogr B Biomed Appl ISSN: 1572-6495