Literature DB >> 7790268

A method for delivering accurate and uniform radiation dosages to the head and neck with asymmetric collimators and a single isocenter.

J W Sohn1, J H Suh, S Pohar.   

Abstract

PURPOSE: To investigate the use of asymmetric collimators and a single isocenter for delivering a uniform, accurate dose of radiation to the head, neck, and supraclavicular lymph nodes. METHODS AND MATERIALS: A linear accelerator with a pair of asymmetric collimators is required for this technique. An isocenter was placed at the junction of the lateral head and neck fields and the anterior supraclavicular field. The asymmetric collimators were set longitudinally, by collimator rotation if necessary. The collimators split the radiation beam to all portals. Dose uniformity was measured at the junction with films in solid-water phantoms.
RESULTS: Film dosimetry showed a uniform dose at the junction without hot or cold regions. A digital display tolerance of +/- 1.0 mm for a field size maintained an acceptable uniform dose (+/- 5% dose variation) at the junction. The single isocenter and asymmetric collimators reduced field setup time by half. No table rotation was required to match fields.
CONCLUSION: The asymmetric collimators lead to easy and accurate patient setup. The absence of the trapezoid effect resulted in the complete coverage of the submandibular and cervical nodes without any hot spots.

Entities:  

Mesh:

Year:  1995        PMID: 7790268     DOI: 10.1016/0360-3016(94)00468-Z

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  Clinical implementation of the AAPM Task Group 36 recommendations on fetal dose from radiotherapy with photon beams: a head and neck irradiation case report.

Authors:  K L Prado; S J Nelson; J J Nuyttens; T E Williams; K N Vanek
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

2.  Target splitting in radiation therapy for lung cancer: further developments and exemplary treatment plans.

Authors:  Karl Wurstbauer; Heinz Deutschmann; Peter Kopp; Florian Merz; Helmut Schöller; Felix Sedlmayer
Journal:  Radiat Oncol       Date:  2009-08-14       Impact factor: 3.481