A 3-D finite element analysis of the natural frequencies of vibration of a stapes prosthesis replacement reconstruction of the middle ear.

K R Williams1, A W Blayney, T H Lesser.   

Abstract

In this work, the natural frequencies of vibration of two different stapes prosthesis replacement reconstructions have been analysed using the finite element method. Prosthesis 1 was constructed of fine stainless steel wire and a Teflon base, while prosthesis 2 was made entirely of Teflon. The results have indicated that generally, the first natural frequency of vibration falls as the prostheses become larger and more bulky. However, the fall in the first natural for prosthesis 1 was modest when compared with that of the isolated tympanic membrane. An important variable influencing frequencies and mode of vibration of the reconstruction was the tightness of fit of the prostheses within the remaining ossicular chain. The tightness of fit in this work was modelled as a range of spring constants applied at the stapes pseudo footplate, together with a series of joint elements at the incus tip/prosthesis head for the Teflon implant. When these spring constraints were below approximately 10 N/mm, the reconstruction exhibited normal mode 1 vibration behaviour, but at larger spring values, an abnormal mode 1 became established resembling the normal mode 2 vibration characteristics. The formation of new geometries and surfaces following reconstruction, introduces new constraints between umbo and stapes footplate particularly at the natural/artificial interfaces. These unnatural constraints may inhibit and modify the natural movements normally occurring at the stapes footplate leading to abnormal modes of vibration.

Entities:  

Mesh:

Year:  1995        PMID: 7788932     DOI: 10.1111/j.1365-2273.1995.tb00009.x

Source DB:  PubMed          Journal:  Clin Otolaryngol Allied Sci        ISSN: 0307-7772


  2 in total

Review 1.  Design and optimization of auditory prostheses using the finite element method: a narrative review.

Authors:  Qianli Cheng; Han Yu; Junpei Liu; Qi Zheng; Yanru Bai; Guangjian Ni
Journal:  Ann Transl Med       Date:  2022-06

2.  Characterization of bone-implant fixation using modal analysis: application to a press-fit implant model.

Authors:  P Swider; G Guérin; Joergen Baas; Kjeld Søballe; Joan E Bechtold
Journal:  J Biomech       Date:  2009-05-22       Impact factor: 2.712

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.