Breast dose during electron-beam CT: measurement with film dosimetry.

C H McCollough1, H H Liu.   

Abstract

PURPOSE: To measure the breast dose and dose distribution from examinations performed with electron-beam computed tomography (CT).
MATERIALS AND METHODS: Radiographic film was placed bilaterally at three depths in the breasts of an anthropomorphic phantom and exposed with clinically relevant protocols. The average breast dose and the dose distribution were obtained from the digitized and calibrated film and compared with those obtained with conventional CT.
RESULTS: Breast doses from electron-beam CT chest, coronary artery calcium, and cardiac function examinations were 21.9, 2.87, and 1.55 mGy, respectively. Within the breast, the dose decreased linearly by a factor of three from the axilla to the sternum. The breast dose from a conventional CT chest examination ranged from 18 to 33 mGy and did not vary substantially across the breast.
CONCLUSION: For chest examinations, the average breast dose from an electron-beam CT scanner was comparable to that from conventional CT scanners, despite the differences in dose distribution.

Entities:  

Mesh:

Year:  1995        PMID: 7784559     DOI: 10.1148/radiology.196.1.7784559

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  Radiation doses in computed tomography. The increasing doses of radiation need to be controlled.

Authors:  M M Rehani; M Berry
Journal:  BMJ       Date:  2000-03-04

2.  Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans.

Authors:  Rafel Tappouni; Bradley Mathers
Journal:  ISRN Radiol       Date:  2012-11-26